User guide
8-2
IM 11Y01B01-01E-A 6th Edition :Feb 13, 2013-00
<8. VALIDATION AND CALIBRATION>
8 .1 Off-Line Manual/Automatic Checking and Off-Line Calibration
To perform either “Ofine Calibrations” or “Ofine Validations”, the Launch unit and Detect unit need to be
removed from their respective process interface and connected to an Off-Line Calibration Cell. If the analyzer is
installed on an extractive ow cell (enhanced ow cell or low volume ow cell) then it is already capable of being
validated/calibrated in off-line mode.
Before removing the units from the process, ensure that the process gas and window purge gas are correctly
isolated to prevent excessive pressure (process or purge gas) from being released or from over-pressuring the
analyzer.
AVOID SHOCK AND IMPACT TO THE ANALYZER
PERMANENT LASER DAMAGE MAY OCCUR
Please note that you can select the block mA value for automatic
Off-Line Zero Calibration, Off-Line Span Calibration, Off-Line
Validations and On-Line Validations:
Figure 59