User guide
3.11.1 Purging Analyzer for Hazardous Areas (with On-Line Validation)
 • NEC/CSA Class 1, Division 2, Groups A-D
 • ATEX Zone 2 CAT 3 Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge control 
system is fi tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pressure switch 
alarm contacts (open on loss of purge pressure). The purging should be carried in sequence typically as shown 
below. All purge gas connections are ¼” od Tube fi ttings.
3.11.2 Purging Analyzer for Hazardous Areas (without On-Line Validation)
 • NEC/CSA Class 1, Division 2, Groups A-D
 • ATEX Zone 2 CAT 3 – Dual regulators must be used on the inlet!
The block diagram below shows the sections of the analyzer that require nitrogen purging. A Z-Type purge 
control system is fi tted the Launch Unit and it includes a local indicator (Bright Green, rugged light) and pres-
sure switch alarm contacts (open on loss of purge pressure).
Figure 26 - Purge Flow Diagram when not using on line validation
Main
Electronic
Housing & 
Purge
System 
Laser
Module
Check Gas 
Flow Cell 
Process 
Interface
Process 
Interface
Detect
Module
Detect
Electronic
Housing
Nitrogen or I/A Purge Gas 
Main 
Electric 
Housing 
& Purge 
System
Detect 
Electronic 
Housing
Main
Electronic
Housing & 
Purge
System 
Laser
Module
Check Gas 
Flow Cell 
Process 
Interface
Process 
Interface
Detect
Module
Detect
Electronic
Housing
Nitrogen or I/A 
Purge Gas 
Nitrogen or I/A 
Purge Gas 
Figure 25 - Purge Flow Diagram when using on line validation
Main 
Electric 
Housing 
& Purge 
System
Detect 
Electronic 
Housing
Nitrogen or I/A Purge
Gas
Check Gas
Flow Cell
<3 INSTALLATION AND WIRING> 3-20