User guide
    <4.  Display and Operations while Connected to GC8000>
4-40
IM 11B06C01-01E
3)  Specify the stream number, analysis start time, and the number of re-integrating 
processes. Subsequently, clicking the “Execute” button performs re-integration of multiple 
chromatograms. At this time, re-integration is performed for the analysis results of the 
analysis count tracked back from the analysis start time.
4)  When re-integration has been completed, the execution result (success or fail) of the 
individual chromatogram re-integration is displayed in the dialog box. Failed re-integration 
results are not updated.
5)  To exit, click the “Close” button. Then, the Re-integrate Multiple Chromatogram dialog box 
closes.
Multiple chromatogram re-integration is determined to fail in the following cases:
•  The number of peak assignments is different from that in the re-integration reference 
chromatogram.
•  Peak detection failed.
•  The concentration value after re-integration falls out of the range.
CAUTION
“Distillation = Yes” is set in the analyzer currently connected or loaded parameter le, re-
integration cannot be performed.
4.4.11  Pen Setting
Select the Set Pen command in the Graph menu to display the Set Pen dialog box.
Figure 4.45  Set Pen dialog box (Example in the right shows a combo box is selected)
From this dialog box, different types of chromatograms to be displayed with pen 1 to 8 can be 
selected.
From a combo box for each pen color, select a chromatogram you want to display. Depending on 
the type of selected chromatogram, a different dialog box may be invoked.
The following shows the graph names that can be selected from the combo box.