User guide
    <4.  Display and Operations while Connected to GC8000>
4-17
IM 11B06C01-01E
4.3.14  Operating Temperature Controllers
To operate a temperature controller, rst check the GCM is in the Manual status. From the 
Manual/Range menu, select the “Oven1” and “Heater1” commands. The Operate Temperature 
Controller dialog box for Oven 1 appears.
Figure 4.21  Operate Temperature Controller dialog box
The Operate Temperature Controller dialog box shows the SV value, PV value, On/Off status and 
temperature control status of Isothermal oven, LSV and FPD. The PV value, On/Off status, and 
temperature control status are refreshed in every second.
To turn the temperature controller Off, just click the ON display. When a conrmation message 
showing “The temperature controller is turned Off” appears in the message box, respond with 
“Yes” and the temperature controller turns Off.
Similarly, when you turn the temperature controller On, click the OFF display. As a conrmation 
message showing “The temperature controller is turned On” appears in the message box, 
respond with “Yes” and the temperature controller turns On.
You can also call the Operate Temperature Controller dialog box by clicking the temperature 
display of the oven.
The dialog box shows the temperature controller status as follows.
Isothermal oven: Under preparation, Constant temperature control
5th Edition : Nov. 16, 2011-00