User guide

<3. Overview Window>
3-15
IM 11B06C01-01E
(3) Enter the le name for the setting parameters to be saved under, preceding the extension
“.px1” (“.pr7”), and click on the “Data” directory which is preserved for saving.
Example: Where “system01” is specied as the le name:
system01. pr7
(4) Click the OK button.
The AS_MAIN dialog box appears, indicating that uploading the setting parameters from the
analyzer is taking place.
F0323.ai
When all the setting parameters have been saved to the le, the FileMan dialog box
appears.
Note that uploading the parameters to the Engineering Terminal takes about 10 seconds.
F0324.ai
(5) Click the OK button.
This completes the uploading process.
HELP !
If you wish to interrupt the uploading of setting parameters,
Click on the Interrupt command in the File menu.
A Conrmation dialog box appears.
Click the Yes button.
n Downloading
To download setting parameters from the Engineering Terminal to an analyzer, do the following:
CAUTION
If the parameters have been written to an analyzer, exit the Engineering Terminal and then restart
it. If you do not restart, data cannot be displayed properly.
5th Edition : Nov. 16, 2011-00