User guide
    <Appendix B  List of Messages>
B-3
IM 11B06C01-01E
n  Analyzer Operation
Message Meaning Countermeasures
[Operation Mode]
Operation command was not 
accepted by the analyzer.
An unauthorized level user is 
attempting to change the operation 
mode or operate a valve.
Only level C users can change the 
operation mode or operate valves.
Valves can only be operated 
in Manual mode.
An attempt is being made to 
operate a valve in a mode other 
than the Manual mode.
To operate a valve, choose the 
Manual mode. Note that only 
level C users can carry out this 
operation.
Detectors can be operated 
only in Manual mode.
An attempt is being made to 
operate a detector in a mode other 
than the Manual mode.
To operate a detector, choose 
the Manual mode. Note that only 
level C users can carry out this 
operation.
Operation pattern, valve, or 
detector cannot be operated 
because the user level is 
inappropriate.
An unauthorized level user is 
attempting to change the operation 
pattern or operate a valve.
Only level B or C users can 
change the operation pattern.
Valves can only be operated by 
level C users.
[Set Up]
The entered data is out of 
range.
An invalid value has been entered 
on Valve/Peak setting dialog box.
Enter the correct value.
The entered data is the same 
as the current data.
“Send” button has been clicked 
when the data is not changed on 
the Valve/Peak setting dialog box.
Click the “Send” button after 
changing the data.
Sending data to analyzer 
failed.
The data transmission has failed 
on the Valve/Peak setting dialog 
box.
Try again after checking the 
operation status of PCAS and 
Analyzers, and connection status 
of the cables.
Please change the LCD mode 
on the GC1000 to remote 
mode.
The Valve/Peak setting is not 
available when the analyzer is 
under “local mode”.
Try again after changing to 
“remote mode” at the analyzer.
A user has changed 
a parameter in the 
GC1000.¥nPlease restart 
Analyzer Operation Window 
by the [Display][Analyzer 
ID] command of Overview 
window.
Due to the parameter change by 
LCD panel at the analyzer, there 
is a descrenpacy of the setting 
parameter information on ASET.
The Valve/Peak setting from ASET 
is not available.
Re-load the initial database 
from the analyzer, then setting is 
available.
n  Chromatogram
Message Meaning Countermeasures
[Time/Pressure 
Display]
Please check Temp. or Press. 
box.
Temperature/Pressure is not 
selected on the Temp/Pressure 
dialog box.
Select at least Temperature or 
Pressure.
Pen select
Select two graphs. 2 graphs are not selected at 
Differential chromatogram display.
Select 2 graphs.
Illegal le type An attempt was made to open a 
le other than a chromatogram le.
Select the correct le.
Already opened The selected differential 
chromatogram has already been 
opened by another pen.
Select another differential 
chromatogram or erase the 
opened differential chromatogram.
n  Alarm
Message Meaning Countermeasures
History 
erasing
Alarm history cannot be reset 
by a level-A user.
An unauthorized level user is 
attempting to reset the alarm 
history.
Only level B or C users can reset 
the alarm history.