    <4.  Display and Operations while Connected to GC8000>
4-61
IM 11B06C01-01E
Re-integration on Individual Peaks 
(select it from a menu that opens with a 
double click on the left mouse button or a 
click on the right mouse button)
Re-integration on Analysis Start Date and Time 
(select it from a menu that opens with a double 
click on the left mouse button or a click on the 
right mouse button)
The re-integrated results are updated in all of the Concentration Analysis, Retention Time, and 
Analysis Results windows. However, if GCM is displayed, only the peak value of the selected 
SYS number is updated.
When the history chromatogram is not saved yet, the re-integration cannot be performed.
l  Re-integration from File Concentration Analysis History and File Retention 
Time History
In online mode, the operation is the same as that for “Re-integration from the Concentration 
Analysis History and Retention Time Analysis History”
However, the analysis results only update the called display window.
Example: If re-integration is performed from the File Concentration Analysis History window, 
only the File Concentration Analysis History window is updated and other windows (such as File 
Retention Time window) are not updated.
In off-line mode, Re-integration (Off-line) is displayed in the right-click menu. When executed, the 
re-integration le select window opens, where you can select the le required for re-integration.
The re-integration results only update the called display window.
Figure4.75 Re-integrationleselectwindow