User guide
    <4.  Display and Operations while Connected to GC8000>
4-57
IM 11B06C01-01E
l  Displaying retention time (history) results from the past
1)  Select Calibration Coefcient History Window in the Window menu, or the Calibration 
Coefcient History Window command button on the toolbar to display the Calibration 
Coefcient History window.
2)  Select SYS Number Select in the Display Item menu, or the system number or GCM of the 
retention time result you want to display from the system number select combo box on the 
toolbar.
3)  Select Open History in the File menu, or the Open History command button on the toolbar to 
open the history setting dialog box.
4)  From the radio button selecting Display from Latest or Display from History, select Display 
from History, then apply it by pressing the OK button.
5)  From the Display from History combo box, select the desired year, month and day, and 
apply it using the OK button.
Figure 4.71  Year and month setting dialog box
6)  The calibration coefcient results are displayed for up to 5,000 past analyses tracing back 
from the specied le. The analysis results are then not automatically updated.
5th Edition : Nov. 16, 2011-00










