User guide
<4. Display and Operations while Connected to GC8000>
4-57
IM 11B06C01-01E
l Displaying retention time (history) results from the past
1) Select Calibration Coefcient History Window in the Window menu, or the Calibration
Coefcient History Window command button on the toolbar to display the Calibration
Coefcient History window.
2) Select SYS Number Select in the Display Item menu, or the system number or GCM of the
retention time result you want to display from the system number select combo box on the
toolbar.
3) Select Open History in the File menu, or the Open History command button on the toolbar to
open the history setting dialog box.
4) From the radio button selecting Display from Latest or Display from History, select Display
from History, then apply it by pressing the OK button.
5) From the Display from History combo box, select the desired year, month and day, and
apply it using the OK button.
Figure 4.71 Year and month setting dialog box
6) The calibration coefcient results are displayed for up to 5,000 past analyses tracing back
from the specied le. The analysis results are then not automatically updated.
5th Edition : Nov. 16, 2011-00