Instruction Manual
  <1. Overview>
1-71
IM 11B08A01-01E
  When all the calibration (or validation) patterns of which automatic execution is enabled are 
completed, the operation mode changes to the stop mode in the stream specifi cation status.
Run
W7 W7 W8
S8S7S7
Time
Stream specification 1 Calibration 1 Stream specification 1
Start of automatic calibration
Pause Stop
  In the case where a stop command is received during the run mode in the (automatic) 
calibration/validation status [4E2]
  The operation mode is immediately prepared to be changed from the run mode to the stop 
mode in the calibration (or validation) status.
  The operation mode changes to the stop mode in the previous measurement status when 
the current measurement is completed.
Run
W7 W7 W8
S7S7
Time
Stream sequence 1 Calibration 1 Stream sequence 1
Start of automatic calibration
Stop Stop
[4E2]
Stop command
  In the case where a forced stop command is received during the run mode in the (manual) 
calibration/validation status [4F2]
  The operation mode is immediately changed to the stop mode in the previous measurement 
status.
  In the case where a forced stop command is received when the operation mode is prepared 
to be changed from the run mode to the stop mode in the (manual) calibration/validation 
status [4F3]
  The operation mode is immediately changed to the stop mode in the previous measurement 
status.
Stop StopRun
(Waiting for stop)
W7 W7
S7S7
Time
Stream sequence 1 Calibration 1 Stream sequence 1
Forced stop command
Start of automatic calibration
[4F2] [4F3]
(Stop command)
[4F3]
2nd Edition : May 11, 2012-00
1