Instruction Manual
<1. Overview>
1-65
IM 11B08A01-01E
Run
W7 W7 W8
W1
S1 S1S8
W1 W1
S1 S7S7
Time
Stream specification 1 (Continuous)
Calibration 1 Stream specification 1 (Continuous)
Assinning calibration 1
In the case where a (semiautomatic) calibration/validation command is received during the
run mode in the stream specifi cation status (for N times of measurement)
Calibration (or validation) is started when the Nth round of the measurement of the currently-
measured stream is completed.
Users at Level C or higher can cancel the calibration/validation command until the
measurement of the currently-measured stream is completed.
When calibration (or validation) is completed for all the streams specifi ed for the selected
calibration (or validation) number, the operation mode changes to the stop mode in the
stream specifi cation status (for N times of measurement).
Run
W7 W7 W8
S8
W1
S1
W1
S1 S7S7
Time
Stream specification 1 (Twice) Calibration 1
Stream specification 1 (Twice)
Assinning calibration 1
Run command
Stop Stop
In the case where a (semiautomatic) calibration/validation command is received during the
stop mode in the stream specifi cation status
Calibration (or validation) is immediately started.
When calibration (or validation) is completed for all the streams specifi ed for the selected
calibration (or validation) number, the operation mode returns to the stop mode in the
stream specifi cation status.
Run
W7 W7 W8
S8S7S7
Time
Stream specification 1 Calibration 1 Stream specification 1
Assinning calibration 1
Stop Stop
In the case where a (semiautomatic) calibration/validation command is received during the
pause mode in the stream specifi cation status
Calibration (or validation) is immediately started.
When calibration (or validation) is completed for all the streams specifi ed for the selected
calibration (or validation) number, the operation mode changes to the stop mode in the
stream sequence status.
Run
W7 W7 W8
S8S7S7
Time
Stream specification 1 Calibration 1 Stream specification 1
Assinning calibration 1
Pause Stop
2nd Edition : May 11, 2012-00
1