Instruction Manual

<1. Overview>
1-59
IM 11B08A01-01E
Run Stop StopRun
W7 W7 W8
S8
W1 W2
S1 S7S7
Time
Stream sequence 1 Calibration 1
(Replaceable) (Replaceable)
Stream sequence 1
Assinning calibration 1
Start of calibration (or validation) End of calibration (or validation)
In the case where a (manual) calibration/validation command is received during the stop
mode in the stream sequence status
The measurement status immediately changes to “Replaceable” in the stop mode in the
requested calibration (or validation) status.
Issue a command to start calibration (or validation) after con rming that manual
replacement of the ow with the sample has been completed. When calibration (or
validation) is completed for all the streams speci ed for the selected calibration (or
validation) number, the operation mode returns to the stop mode in the stream sequence
status.
Stop StopRun
W7 W7 W8
S8S7S7
Time
Stream sequence 1 Calibration 1
(Replaceable) (Replaceable)
Stream sequence 1
Assinning calibration 1 Calibration (or validation) start command Calibration (or validation) stop command
In the case where a (manual) calibration/validation command is received during the pause
mode in the stream sequence status
The measurement status immediately changes to “Replaceable” in the stop mode in the
requested calibration (or validation) status.
Issue a command to start calibration (or validation) after con rming that manual
replacement of the ow with the sample has been completed. When calibration (or
validation) is completed for all the streams speci ed for the selected calibration (or
validation) number, the operation mode changes to the stop mode in the stream sequence
status.
Pause Stop StopRun
W7 W7 W8
S8S7S7
Time
Stream sequence 1 Calibration 1
(Replaceable) (Replaceable)
Stream sequence 1
Assinning calibration 1 Calibration (or validation) start command Calibration (or validation) stop command
In the case where a (manual) calibration/validation command is received during the run
mode in the stream speci cation status
The measurement status is immediately prepared to be changed to the calibration (or
validation) status.
In the case of the stream speci cation status (for continuous measurement), the
measurement status immediately changes to “Replaceable” in the stop mode in the
requested calibration (or validation) status.
In the case of the stream speci cation status (for N times of measurement), the
measurement status changes to “Replaceable” in the stop mode in the requested calibration
(or validation) status when the Nth round of the measurement of the currently-measured
stream is completed.