Figure 1.27  An operation example of the calibration (validation) stop window on the GCM operation 
status screen on the GC-HMI analyzer operation display
  In the case where a (manual) calibration/validation command is received during the run 
mode in the stream sequence status
  The measurement status is immediately prepared to be changed to the calibration (or 
validation) status.
  The measurement status changes to “Replaceable” in the stop mode in the calibration (or 
validation) status when the measurement of the currently-measured stream is completed.
  Users at Level C or higher can cancel the calibration/validation command until the operation 
mode changes to the stop mode.
  Issue a command to start calibration (or validation) after confi rming that manual 
replacement of the fl ow with the sample has been completed. When calibration (or 
validation) is completed for all the streams specifi ed for the selected calibration (or 
validation) number, the operation mode returns to the stop mode in the stream sequence 
status.