Instruction Manual
  <1. Overview>
1-39
IM 11B08A01-01E
Run Pause Stop Run Pause Stop
Run Stop
Stream sequence
Stream designation
Calibration/validation
*1
*1*1*1*1
*2
*1
(1) (2) (1) (2)
*1
(3) (4)
(A) (A) (A)(A)(B) (B) (B)(B)
*1:  Command can be canceled.
*2:  Only for manual
•  If a calibration/validation command is issued during the pause mode or the stop mode in the stream sequence or stream specifi ation 
(N times or continuous (set to “0”)) status
  (A) The measurement status and the operation mode innediately changes to the requested calibration/validation status and the 
run mode, respectively.
  (B) The measurement status and the operation mode respectivily return to the stream sequenace status and the stop mode 
when calibration/validation is complete for all the spwcifi ed streams.
•  If a calibration/validation command is issued during the run mode in the stream sequence status
  (1) The measurement status changes to the requested calibration/validation status when the measurement of the currently-
measured stream is completed.
  (2) The measurement status and the operation mode respectively return to the stream sequence status and the run mode when 
calibration/validation is completed for all the specifi ed streams. The suspended measurement is (not reset but) resumed from 
the next stream.
•  If a calibration/validation commands is issued during the run mode in the stream specifi cation (N times) status
  (3) The measurement status changes to the requested calibration/validation status when the Nth round of the spwcifi ed 
measurement is completes.
  (4) The measurement status and the operation mode respectively return to the stream specifi cation (N times) status and the stop 
mode when calibration/validation is completed for all the specifi ed streams.
•  If a calibration/validation command is issued in the stream specifi cation (Continuous (set to “0”))status
  (3) The measurement status changes to the requested calibration/validation status when the measurement of the currently-
measured stream is completed.
  (4) The measurement status and the operation mode respectively return to the stream specifi cation (Continuous) status and the 
run mode when calibration/validation is completed for all the specifi ed streams.
Figure 1.19  Transition among measurement statuses and operation modes
(1) Operation  description
For example, this section describes the operation where the stream sequence, calibration, and 
validation are set as shown in Table 1.5, 1.6, and 1.7, respectively.
Table 1.5
1st stream 2nd stream 3rd stream
Stream sequence 1 Stream 1 Stream 2 Stream 3
Stream sequence 2 Stream 4 Stream 5 Stream 6
Table 1.6
Calibration 
stream
Measurement 
frequency
Automatic 
calibration
Validation 
stream 
number 
before 
calibration
Measurement 
frequency
Validating 
stream 
number 
after 
calibration
Measurement 
frequency
Calibration 1 Stream 7 2 Valid 0 0 Stream 8 1
2nd Edition : May 11, 2012-00