Instruction Manual
    <Appendix G User Program>
App.G-13
IM 11B08A01-01E
GET_C_END_T Command
Function  Acquires analysis results of specifi c peak during calculation (stop time).
Format GET_C_END_T(N)
  N: Peak number (1 to 999)
  Return value: Single precision real number type
Explanation Omitted
GET_C_HEIGHT Command
Function  Acquires analysis results of specifi c peak during calculation (height).
Format GET_C_HEIGHT(N)
  N: Peak number (1 to 999)
  Return value: Single precision real number type
Explanation Omitted
GET_C_RT Command
Function  Acquires analysis results of specifi c peak during calculation (retention time).
Format GET_C_RT(N)
  N: Peak number (1 to 999)
  Return value: Single precision real number type
Explanation Omitted
GET_C_START_H Command
Function  Acquires analysis results of specifi c peak during calculation (peak start level).
Format GETC_START_H(N)
  N: Peak number (1 to 999)
  Return value: Single precision real number type
Explanation Omitted
GET_C_START_T Command
Function  Acquires analysis results of specifi c peak during calculation (peak start time).
Format GET_C_START_T(N)
  N: Peak number (1 to 999)
  Return value: Single precision real number type
Explanation Omitted
GET_DHM Command
Function  Acquires the current date, hour and minute and assigns acquired results into 
DATE, HOUR and MIN.
Format GET_DHM
  Return value: None
Explanation  Used only with GCUD.
GET_GC_COMM_STS Command
Function  Acquires the MODBUS communication status with the specifi ed analyzer ID.
Format GET_GC_COMM_STS(GC_ID)
  GC_ID: Analyzer ID (1 to 240)
  Return value: Integer
  Return 0 for communication failure, 1 for in communication, –1 for unregistered. 
Explanation Omitted
GET_S_AREA Command
Function  Acquires the latest analysis results of specifi ed stream and peak (area).
Format GET_S_AREA(M, N)
  M: Stream number (1 to 31)
2nd Edition : May 11, 2012-00