Instruction Manual
<Appendix G User Program>
App.G-11
IM 11B08A01-01E
9 GET_GC_COMM_STS
10 GET_S_AREA
11 GET_S_HEIGHT
12 GET_S_CALM_STS
13 GET_S_CONC
14 GET_S_END_H
15 GET_S_END_T
16 GET_S_RT
17 GET_S_START_H
18 GET_S_START_T
19 MSS_CAL
20 MSS_STR
21 MSS_STR_SEQ
22 MSS_VAL
23 OMC_PAUSE
24 OMC_STOP
25 RANGE_S
26 READ_AI
27 READ_DI
28 READ_DO
29 READ_GC_AI
30 READ_GC_DI
31 READ_NV_DBL
32 READ_NV_INT
33 READ_NV_SNG
34 WRITE_NV_DBL
35 WRTIE_NV_SNG
36 WRITE_NV_INT
37 OMC_RUN Script for fi xed period
38 WAIT
39 GET_C_AREA Script for peak detection stop time
40 GET_C_CALM_STS
41 GET_C_CONC
42 GET_C_END_H
43 GET_C_END_T
44 GET_C_HEIGHT
45 GET_C_RT
46 GET_C_START_H
47 GET_C_START_T
48 GET_STR_NO
49 PUT_C_CONC
Regarding to command content, see the following for details.
DO_OFF Command
Function Switches off the specifi ed DO number.
Format DO_OFF(N)
N: DO number (1 to 32)
Return value: None
Explanation Omitted
DO_ON Command
Function Switches on the specifi ed DO number.
Format DO_ON(N)
N: DO number (1 to 32)
Return value: None
Explanation Omitted
GET_ALARM_1 Command
Function Acquires level 1 alarm status.
Format GET_ALARM_1
Return value: Integer
Return 1 for level 1 alarm, 0 if other.
Explanation Omitted
GET_ALARM_2 Command
Function Acquires level 2 alarm status.
Format GET_ALARM_2
Return value: Integer
2nd Edition : May 11, 2012-00
App.G