Instruction Manual

<Appendix E Computation Scheme>
App.E-13
IM 11B08A01-01E
The computation formula is:
c
i
=
(f
j
s
j
)
f
i
s
i
k
j=l
Where:
c
i
: concentration of peak i
s
i
: area of peak i
f
i
: sensitivity correction coefficient of peak i
(f) Total Compensation
Total compensation computes the concentration measured in a stream so that the total value of
the measurement components is the total value speci ed in each stream.
The computation formula is:
c
ni
= c
total
c
j
c
i
k
j=l
Where:
c
ni
: result of total compensation for peak i (excluding the concentrations
of the “Total compensation disable” peaks)
c
ni
: concentration of peak i
c
total
: total of the concentration values
(g) Variation Coef cient Computation
Variation coef cient computation computes the variation coef cient for concentration.
The computation formula is:
c
avrg
=
n
c
i
c
v
=
1
c
span
(c
i
–c
avrg
)
2
n
Where:
c
i
: concentration of peak i
c
span
: measurement span of peak i
c
v
: variation coefficient of concentration for peak i
(3) Deviation Component Computation
Deviation component computation computes the difference between the concentration values
measured in a stream and the total concentration speci ed for each stream.
Deviation coef cient computation can be made for obtained concentration values. See the
section on variation coef cient computation in peak processing.
If the result of a computation is negative, the result is counted as 0.
2nd Edition : May 11, 2012-00
App.E