Coexistence of slope/time method with regular peak detection actions
Like the slope gating, time gating and zone gating method, the slope/time method is
performed by setting at the peak unit. Therefore, the slope/time method is considered as
one of the regular detection actions, allowing it to coexist with other detection methods in
one stream.
(g) Time/slope Method
Unlike the usual slope detection method or time gating method, time/slope method performs a
peak detection by using the time gating method for the peak on detection and the slope detection
method for the peak off.
Coexistence of time/slope method with regular peak detection actions
Like the slope gating, time gating and zone gating method, the time/slope method is
performed by setting at the peak unit. Therefore, the time/slope method is considered as
one of the regular detection actions, allowing it to coexist with other detection methods in
one stream.
Peak Waveform processing
This is the same as in peak waveform processing in Manual mode. (See Section 2.3.2.)
However, area calculation by tangent skimming or vertical separation can be speci ed for each
peak.
Concentration Computation Processing
Concentration is computed when level processing for the last peak is completed in each stream.
Corrected area normalization or the standard method (external standard of indirect method) can
be group-speci ed in the stream common list.
(a) External Standard (Approximation by Cubic Equation)
The external standard (approximation by cubic equation) computes concentration from the area
value based on the prede ned coef cient.
The speci ed standard peak is used for the coef cient , S
b
, R, K, A and B.
The computation formula is:
C = RK(S
i
/S
b
) (A (S
i
/S
b
)
2
+ B(S
i
/S
b
) + 1)
Where:
C : concentration
: calibration factor
S
b
: standard area (or height)
S
i
: measured area (or height)
R : measurement range
K : coefficient
A : coefficient
B : coefficient
(b) External Standard (Approximation by Linear Equation)
The external standard (approximation by linear equation) calculates a coef cient of the linear
equation from the area (or height) of two peaks speci ed as standard peaks speci ed as
standard peaks and calculates the concentration from the area (or height) measured using the
coef cient.
The equations are as follows.
(1) During calibration
During calibration, if the range of the calibration coef cient is zero, calculate the standard
area, if speci ed, calculates the calibration coef cient.
2nd Edition : May 11, 2012-00