Instruction Manual

<Appendix E Computation Scheme>
App.E-7
IM 11B08A01-01E
Actions:
According to the change in the retention time of each peak that is speci ed as standard, the
gate time of the speci ed peak and processing time of the speci ed signal level are changed.
When only one standard peak is speci ed, the gate time and processing time are obtained in
proportion with the retention time. When two or more standard peaks are speci ed, the gate time
and processing time are obtained for each peak by performing a linear approximation on the rst
peak and the last peak.
The variation coef cient of the obtained retention time is shared by all peaks.
F0205020102.ai
Initial chromatogram
Initial gate
Initial gate
Initial chromatogram
Chromatogram after change
Gate after tracking processing
Standard peak
(c) Time Gating
Time gating processing detects peaks within the prede ned time interval when the following
condition is satis ed:
The detector signal value is higher than the detection level within the range between the speci ed
gate-on time and gate-off time.
The detection level can be set for each peak based on the signal level upon analysis start (or the
base level when the base level setting is speci ed).
F0205020103.ai
Base level setting
Detection level
Detection level
Base level
Base level
Peak Peak
GateGate
(d) Slope Gating
Slope gating processing detects peaks based on the change in the time of detector signal
data. The time when detection is started is prede ned as the gate time. When all the following
conditions are satis ed, it is assumed that the peak is started (on):