Instruction Manual
    <Appendix E Computation Scheme>
App.E-6
IM 11B08A01-01E
Peak area/peak height
F020502_2.ai
Select one of the following Concentration computation processing
External standard 
(approximation 
from cubic 
equation)
External standard 
(approximation 
from linear 
equation)
Total 
compensation 
operation
Variation 
coefficient
External standard 
(calibration 
computation)
Indirect method
Corrected area 
normalization
Select wheter to execute
Concentration value Variation coefficient
 Peak Detection
The details on square processing, retention time detection processing, and tangent processing 
are the same as in peak processing in Manual mode.
(a)  Base Line Correction Processing
 Base Line Correction
In base line correction, stored chromatogram data are subtracted from the measured detector 
signal data, and processing following the peak detection are carried out on the subtracted data. 
When storage is specifi ed, base line correction is not done.
F02040101.ai
Measured chromatogram
Stored chromatogram
Corrected chromatogram
You can select whether or not correction is carried out for each stream.
(b)  Auto-tracking Peak Processing
As the retention time changes, gates for each peak are displaced. To prevent this displacement, 
auto-tracking peak processing automatically changes the gate time for each peak according to 
the change in the retention time.
2nd Edition : May 11, 2012-00