Instruction Manual

<Appendix D Action of External Input and Output Signals>
App.D-27
IM 11B08A01-01E
Answerback (command not executed) (#W or #W#)
1234 When receiving an invalid-format command 1 2345
#W
CR LF
#W#
CR LF
TIP
Answerback (comannd not executed) #W and #W#
#W: When the command includes an out-of-range parameter
When sending a message to another task has failed
#W#: Invalid format
Invalid command length
Unde ned command
Invalid analyzer ID for the GC1000/GC8
Input instruction (DCS to GC8000)
Data transmission request (#A)
1234
#A
CR LF
Data retransmission request (#R)
1234
#R
CR LF
Example of communication “with procedure”
The GC8000 sends “Input command prompt (#T)” at an interval of 8 seconds.
<When any data of analysis value, calibration factor or alarm is communicated.>
In the case of “with procedure” and upon outputting analysis value (concentration/calculation),
calibration factor, or alarm data, the GC8000 issues the relevant data packet to DCS by the
following procedure.
Normal comminucation
(1) When the GC8000 is ready to send data to DCS, the GC8000 rst sends “Data receiving
request (#E)” to DCS.
(2) Upon receiving “Data transmission request (#A)” from DCS, the GC8000 sends all data.
After the completion, the GC8000 sends “Data transmission completion (#F).”
*1
*2 *2
DCS #A #A - - - - #A
GC8000 #E Data 1 - - - - - Data n #F
*1: 3 seconds or less, *2: 0.04 seconds
When not receiving “Data transmission request (#A or #R)”
If the GC8000 does not receive “Data transmission request (#A or #R)” from DCS for about
3 seconds after sending “Data receiving request (#E)”, the GC8000 sends “Data receiving
request (#E)” again. The GC8000 repeats the same request up to six times (7 times in total)
until the GC8000 receives a reply from DCS. If it does not receive any reply, the GC8000
sends “Data transmission completion (#F)” to DCS.
*1 *1
DCS
GC8000 #E #E #E #E #E #E #E #F
*1: 3 seconds
2nd Edition : May 11, 2012-00
App.D