Instruction Manual

<Appendix D Action of External Input and Output Signals>
App.D-25
IM 11B08A01-01E
Calibration command (CM/CE/CA)
Calibration/Validation command is executed. Calibrations 4 to 6 and Validations 4 to 6 are not
supported.
1234567 For GC6 12345
CM
CR LF
CA
CR LF
  
ab a
123456789
CE
CR LF


ab
a: Calibration/validation number (1 to 3 = Calibrations 1 to 3; 4 to 6 = Validations 1 to 3)
b: Analyzer number (1 to 9 when M comes in the second digit, and 1 to 240 or 001 to 240 when E comes in the second digit)
Stream sequence change command (UM/UE)
Stream sequence change command is executed.
1234567
UM
CR LF
 
ab
123456789
UE
CR LF


ab
a: Stream Sequence number (1 to 8)
b: Analyzer number (1 to 9 when M comes in the second digit, and 1 to 240 or 001 to 240 when E comes in the second digit)
Start command (BM/BE)
Run command is executed.
12345 For GC6 1234
BM
CR LF
BE
CR LF

a
1234567
BE
CR LF

a
a: Analyzer number (1 to 9 when M comes in the second digit, and 1 to 240 or 001 to 240 when E comes in the second digit)
Stop command (FM/FE/FI)
Stop command is executed.
12345 For GC6 1234
FM
CR LF
FI
CR LF

a
1234567
FE
CR LF

a
a: Analyzer number (1 to 9 when M comes in the second digit, and 1 to 240 or 001 to 240 when E comes in the second digit)
2nd Edition : May 11, 2012-00
App.D