Instruction Manual

<Appendix D Action of External Input and Output Signals>
App.D-23
IM 11B08A01-01E
Type B
<GC1000/GC8>
123456789101112131415161718192021222324252627282930313233343536373839404142434445
D
S
1
2
,,
CR LF
   
ab c d
a: Stream number (01 to 31)
b: Peak number (01 to 99) *Relative peak number
To address all 255 peak numbers, S comes in the second digit (from the head) for the peak numbers 01 to 99, 1 for the numbers
100 to 199, and 2 for the numbers 200 to 255.
c: Analysis value (5 digits)
The data format for analysis values depends on the measurement range for the peak setting (see Table 11).
Analysis values are rounded.
d: Analyzer number (0 to 240) (zero suppression provided)
<GC6>
The Analyzer number (d) to be used in the <GC1000/GC8> format is not provided. Only S comes
in the second digit because the peak numbers are 01 to 99. See Type A.
Type C
If a unit other than ppm or % or no unit is used, data is not output.
<GC1000/GC8>
123456789101112131415161718192021222324252627282930313233343536373839404142434445
D
S
1
2
,,
P
%
PM
,
A:CHL
,
A:CLL
CR LF
     
ab c d e f
a: Stream number (01 to 31)
b: Peak number (01 to 99) *Relative peak number
To address all 255 peak numbers, S comes in the second digit (from the head) for the peak numbers 01 to 99, 1 for the numbers
100 to 199, and 2 for the numbers 200 to 255.
c: Analysis value (5 digits)
The data format for analysis values depends on the measurement range for the peak setting (see Table 11).
Analysis values are rounded.
d: Unit (PPM, %)
e: Concentration error alarm (A:CHL = Higher limit, A:CLL = Lower limit)
They will be dummy spaces if no error occurs.
f: Analyzer number (1 to 240) (zero suppression provided)
<GC6>
The Analyzer number (f) to be used in the <GC1000/GC8> format is not provided. Only S comes
in the second digit because the peak numbers are 01 to 99. See Type A.
(2) Calibration factor data ( xed to 45 characters)
The data is output when Execution is set to “Executed” on the Peak Setup-Speci c screen.
When a CAL (calibration out of range) or RPT (calibration repeatability error) alarm occurs, this is
not output.
<GC1000/GC8>
123456789101112131415161718192021222324252627282930313233343536373839404142434445
CC , , .
,COE
CR LF

   
abcd e
a: Calibration number (1 to 3)
b: Peak number (01 to 99) *Relative peak number
c: Calibration factor (5 digits)
Calibration factors are rounded. The position of the decimal point is xed.
d: Calibration factor error (COE)
They will be dummy spaces if no error occurs.
e: Analyzer number (1 to 240) (zero suppression provided)
2nd Edition : May 11, 2012-00
App.D