Instruction Manual
<Appendix D Action of External Input and Output Signals>
App.D-19
IM 11B08A01-01E
The Calibration/Validation function of D/I is available for semiautomatic calibration. To enable the
function, the relevant calibration or validation number is required to be specifi ed in advance.
11/11/22
15:15:45
D/I Setup
Slot # 1 DI # 1
> - Signal name G1-CAL1
- Process assign Cal/Val
- GCM number 1
- Cal/Val number Calibration 1
Menu Status Slot# DI#
F1 F2 F3 F4 F5 F6
Figure 26
Process Specifi cation: Calibration/Validation (D/I Setting Screen on GC-HMI EtherLCD)
11/11/22
15:15:45
Cal/Val Setup (Main)
GCM # 1
> - Cal(Val) method Semi-auto
- Auto start date 2011/11/22
- Auto start time 17:00
- Time interval 001 day:00:05
Auto cal status Stopping
Menu Start Stop Cal Val GCM#
F1 F2 F3 F4 F5 F6
11/11/22
15:15:45
Calibration Setup
GCM # 1 Cal # 1
> - Cal stream 4
- Cal times 3
- Auto cal Executed
- ValStr# before Cal 3
- ValStr# after Cal 4
- Val before Cal times 1
- Val after Cal times 1
Menu Peak Main GCM# Cal#
F1 F2 F3 F4 F5 F6
Figure 27
Calibration/Validation Setting Screen and Calibration Screen on GC-HMI EtherLCD.
Operation Mode
The Operation Mode function of D/I is a function to give a command to change the operation
mode to a specifi ed operation mode for a specifi ed GCM using contact input as a trigger.
2nd Edition : May 11, 2012-00
App.D