Instruction Manual
    <Appendix D Action of External Input and Output Signals>
App.D-19
IM 11B08A01-01E
The Calibration/Validation function of D/I is available for semiautomatic calibration. To enable the 
function, the relevant calibration or validation number is required to be specifi ed in advance.
11/11/22
15:15:45
D/I Setup 
Slot # 1  DI # 1
> - Signal name  G1-CAL1
 - Process assign  Cal/Val
 - GCM number  1
 - Cal/Val number  Calibration 1
Menu Status Slot# DI#
F1 F2 F3 F4 F5 F6
Figure 26 
Process Specifi cation: Calibration/Validation (D/I Setting Screen on GC-HMI EtherLCD)
11/11/22
15:15:45
Cal/Val Setup (Main) 
GCM # 1 
> -  Cal(Val)  method  Semi-auto
 - Auto start date  2011/11/22
 - Auto start time  17:00
 - Time interval  001 day:00:05
Auto cal status Stopping
Menu Start Stop Cal Val GCM#
F1 F2 F3 F4 F5 F6
11/11/22
15:15:45
Calibration Setup 
GCM # 1  Cal # 1
> - Cal stream  4
 - Cal times  3
 - Auto cal  Executed
 - ValStr# before Cal  3
 - ValStr# after Cal  4
 - Val before Cal times  1
 - Val after Cal times  1
Menu Peak Main GCM# Cal#
F1 F2 F3 F4 F5 F6
Figure 27 
Calibration/Validation Setting Screen and Calibration Screen on GC-HMI EtherLCD. 
 Operation Mode
The Operation Mode function of D/I is a function to give a command to change the operation 
mode to a specifi ed operation mode for a specifi ed GCM using contact input as a trigger.
2nd Edition : May 11, 2012-00
App.D