Instruction Manual
    <Appendix D Action of External Input and Output Signals>
App.D-18
IM 11B08A01-01E
11/11/22
15:15:45
D/I Setup 
Slot # 1  DI # 1
> - Signal name  S1P1-R1
 - Process assign  Range Select
 - Stream number  1  (GCM 1)
 - Peak number  1
 - Range number  1
Menu Status Slot# DI#
F1 F2 F3 F4 F5 F6
Figure 24 
Process Specifi cation: Range Switching (D/I Setting Screen on GC-HMI EtherLCD)
11/11/22
15:15:45
Peak Setup-Specifi c  (2/5)
Stream #1  Peak #1  External 3rd
 - Detector number  1-1
> - Measuring unit  1
 - Measuring range  100.000
 - Gate cut method  Slope gate
 - Integ method  Skimming
 - Peak polarity  Positive
 - Detected slope  0.0010 mV/s
Menu Copy1 Copy2 Strm# Peak#
F1 F2 F3 F4 F5 F6
11/11/22
15:15:45
Multirange Setup 
Range # 1
> - Process (Type)  Peak
 - Process (Detail)  External 3rd
 - Measuring unit  1
 - Measuring range  50.000
 - Ref peak number  #40 #001
Menu Get Range#
F1 F2 F3 F4 F5 F6
Figure 25 
Individual Peak Setting Screen and Multi-Range Setting Screen on GC-HMI EtherLCD
 Calibration/Validation
The Calibration/Validation function of D/I is a function to give a command to change the 
measurement status to the status of a specifi ed calibration or validation number using contact 
input as a trigger.
Six patterns of calibration setting and validation setting can be used for each GCM.
To give a command to all GCMs, set “0” for the GCM number.
2nd Edition : May 11, 2012-00