<Appendix D Action of External Input and Output Signals>
App.D-18
IM 11B08A01-01E
11/11/22
15:15:45
D/I Setup
Slot # 1 DI # 1
> - Signal name S1P1-R1
- Process assign Range Select
- Stream number 1 (GCM 1)
- Peak number 1
- Range number 1
Menu Status Slot# DI#
F1 F2 F3 F4 F5 F6
Figure 24
Process Speci cation: Range Switching (D/I Setting Screen on GC-HMI EtherLCD)
11/11/22
15:15:45
Peak Setup-Speci c (2/5)
Stream #1 Peak #1 External 3rd
- Detector number 1-1
> - Measuring unit 1
- Measuring range 100.000
- Gate cut method Slope gate
- Integ method Skimming
- Peak polarity Positive
- Detected slope 0.0010 mV/s
Menu Copy1 Copy2 Strm# Peak#
F1 F2 F3 F4 F5 F6
11/11/22
15:15:45
Multirange Setup
Range # 1
> - Process (Type) Peak
- Process (Detail) External 3rd
- Measuring unit 1
- Measuring range 50.000
- Ref peak number #40 #001
Menu Get Range#
F1 F2 F3 F4 F5 F6
Figure 25
Individual Peak Setting Screen and Multi-Range Setting Screen on GC-HMI EtherLCD
Calibration/Validation
The Calibration/Validation function of D/I is a function to give a command to change the
measurement status to the status of a speci ed calibration or validation number using contact
input as a trigger.
Six patterns of calibration setting and validation setting can be used for each GCM.
To give a command to all GCMs, set “0” for the GCM number.
2nd Edition : May 11, 2012-00