Instruction Manual
    <Appendix D Action of External Input and Output Signals>
App.D-15
IM 11B08A01-01E
The following four types of measurement (check items) can be set as a trigger of a Component 
alarm. One check item can be set for each contact.
• Concentration
• Retention time
• Variation coeffi cient
• Tailing factor
11/11/22
15:15:45
Alarm Setup 
GCM # 1   Set # 1
> - Stream number  1
 Peak number  1
 - Check item  Conc
 - Maximum  90.000
 - Minimum  100.000
Menu GCM# Set#
F1 F2 F3 F4 F5 F6
Figure 19 
Alarm Setting Screen on GC-HMI EtherLCD 
 Timing
The Timing function of D/O is a function to send information about the time of completion of peak 
detection to the upper systems.
This is used in combination with analog hold output (analysis result output) that uses A/O.
The contact output is turned On and Off at the time set for the D/O operating time in the SYS 
method.
 Calibration/Validation
The Calibration/Validation function of D/O is a function to send out information that a calibration 
or validation number set for each GCM is being executed. The contact output is turned On when 
the measurement status of the GCM set on the D/O setting screen is the status of the specifi ed 
calibration or validation number.
Multiple calibration or validation numbers can be set for one contact.
11/11/22
15:15:45
D/O Setup  (1/2)
11/11/22
15:15:45
D/O Setup  (2/2)
Slot # 1  DO # 1 Slot # 1  DO # 1
> - Process assign  Cal/Val > - Validation 1  On
 - GCM number  1  - Validation 2  On
 - Calibration 1  On  - Validation 3  On
 - Calibration 2  On  - Validation 4  On
 - Calibration 3  On  - Validation 5  On
 - Calibration 4  On  - Validation 6  On
 - Calibration 5  On
 - Calibration 6  On
Menu Status Slot# DO# Menu Status Slot# DO#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 20 
Process Specifi cation: Calibration/Validation (D/O Setting Screen on GC-HMI EtherLCD)
2nd Edition : May 11, 2012-00