Instruction Manual
    <Appendix D Action of External Input and Output Signals>
App.D-13
IM 11B08A01-01E
Table 11
In the status of stream 
sequence 1 or stream 
specifi cation 1
Stop to Pause
Turned On when the mode changes to the pause 
mode
Stop to Run
Turned On when the mode changes to the run 
mode
Run to Pause Kept turned On
Run to Stop
Turned Off when the mode changes to the stop 
mode
Run mode of stream sequence 1* to Run mode of 
stream sequence 2
Turned Off when analysis starts for the stream 2
Pause mode of stream sequence 1* to Stop mode 
of stream sequence 2
Turned Off when the mode changes to the stop 
mode of stream 2
Stop mode of stream sequence 1* to Stop mode of 
stream sequence 2
Run mode of stream sequence 2 to Run mode of 
stream sequence 1*
Turned On when analysis starts for the stream 1
Pause mode of stream sequence 2 to Stop mode 
of stream sequence 1*
Kept turned Off
Stop mode of stream sequence 2 to Stop mode in 
the status of stream sequence 1*
Run mode of stream sequence 1 to Run mode of 
stream specifi cation 1
Kept turned On
Pause mode of stream sequence 1 to Stop mode 
of stream specifi cation 1
Turned Off when the mode changes to the stop 
mode of stream 1
Stop mode of stream sequence 1 to Stop mode of 
stream specifi cation 1
Kept turned Off
Run mode of stream sequence 1* to Calibration/
Validation
Turned Off when analysis starts for the calibration/
validation stream
Pause mode of stream sequence 1* to Calibration/
Validation
Turned Off when analysis starts for the calibration/
validation stream
Stop mode of stream sequence 1* to Calibration/
Validation
Kept turned Off
*  Same for the case of stream specifi cation 1 
11/11/22
15:15:45
D/O Setup 
Slot # 1     DO # 1
> - Process assign  Stream
 - Stream 1 (GCM 1)
Menu Status Slot# DO#
F1 F2 F3 F4 F5 F6
Figure 16 
Process Specifi cation: Stream (D/O Setting Screen on GC-HMI EtherLCD)
 Operation Mode
The Operation Mode function of D/O is a function to send out information that the operation mode 
of each GCM is in the specifi ed mode. The contact output is turned On in the operation mode set 
on the D/O setting screen.
One among the following four types of operation mode is set for each contact point.
• Run
•  Run (Warming up time)
• Stop
• Pause
2nd Edition : May 11, 2012-00