Instruction Manual
    <Appendix D Action of External Input and Output Signals>
App.D-10
IM 11B08A01-01E
 Analog Input
By using optional A/I cards (up to four cards, four inputs per card), up to 16 inputs can be output 
as A/O.
A/I cards can acquire analog information from fi eld instruments (process pressure gauges and 
fl ow meters) and analyzers, which can be calculated with the user program, a new function of the 
GC8000.
11/11/22
15:15:45
Peak Setup-Specifi c  (1/2)
11/11/22
15:15:45
Peak Setup-Specifi c  (2/2)
Stream #1  Peak #1 Stream #1  Peak #1  AI signal 1
 Using/Assigned peak num  8  / 10 > - 0% value  200.000
> - Process (Type)  AI  - 100% value  400.000
 - Process (Detail)  AI signal 1  - Action time  100.0 s
 - Name  Pressure  - SYS number  1
 - Execution  Excuted
 - Synchronize to  #01  #001
 - Output fl ag  Provided (31001)
 - Range number  0
 - Auto gain value  0
Menu Copy1 Copy2 Strm# Peak# Menu Copy1 Copy2 Strm# Peak#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 12 
AI Processing (Individual Peak Setting Screen on GC-HMI EtherLCD)
11/11/22
15:15:45
Multirange Setup 
Range # 1
> - Process (Type)  AI
 - Process (Detail)  AI signal 1
 - 0% value  200.000
 - 100% value  400.000
Menu Get Range#
F1 F2 F3 F4 F5 F6
Figure 13 
AI Processing (Multi-Range Setting Screen on GC-HMI EtherLCD)
11/11/22
15:15:45
A/I Setup 
Slot # 1  AI # 1
> - Signal name  Pressure
 - Filter const  1.000
Menu Status Slot# AI#
F1 F2 F3 F4 F5 F6
Figure 14  A/I Setting Screen on GC-HMI EtherLCD 
2nd Edition : May 11, 2012-00