Instruction Manual

<Appendix D Action of External Input and Output Signals>
App.D-4
IM 11B08A01-01E
Table 3 Example of Stream Identi cation Signal Settings for One GCM with Six Streams
GCM1
Stream 1 Stream 2 Stream 3 Stream 4 Stream 5 Stream 6
STR1 STR2 VAL1 VAL2 CAL1 CAL2
Measurement Measurement Validation Validation Calibration Calibration
GCM1 SLOT4 D/O1 Stream
identi cation
ag
Stream
identi cation
ag
Stream
identi cation
ag
Stream
identi cation
ag
Stream
identi cation
ag
Stream
identi cation
ag
D/O2 1 0 1 0 1 0
D/O3 0 1 1 0 0 1
D/O4 0 0 0 1 1 1
Stream valve number 1 2 3 4 3 4
* Select Up to 7 streams and Starting D/O number for stream identi cation: 4-1 for GCM1. Use one D/O card (5 ch).
* The starting stream valve number and number of stream valves of GCM1 are “1” and “6,” respectively.
* The above settings are for the case where output exists for calibration (or validation) streams.
In the case where output dose not exists for calibration (or validation), the stream identi cation signal of the latest measured
stream other than calibration or validation streams is held.
11/11/22
15:15:45
GCM Setup (2/2)
GCM # 1
> - ATM valve number 1-1
Strm identifying Up to 7 streams
-
1st num StrIdentify DO
Slot #4 -#1
- 1st num StrValve 1
- Stream valve num 6
- Distillation None
Menu GCM#
F1 F2 F3 F4 F5 F6
Figure 4
Stream Identi cation Signal Output: Exist (GCM Setting Screen on GC-HMI EtherLCD)
Table 4 Example of Stream Identi cation Signal Settings for Two GCMs with Three Streams Each
GCM1 GCM2
Stream 1 Stream 2 Stream 3 Stream 4 Stream 5 Stream 6
STR1 VAL1 CAL1 STR2 VAL2 CAL2
Measurement Validation Calibration Measurement Validation Calibration
GCM1 SLOT4 D/O1 Stream
identi cation
ag
Stream
identi cation
ag
Stream
identi cation
ag
D/O2 1 0 1
D/O3 0 1 1
D/O4 0 0 0
GCM2 SLOT5 D/O1 Stream
identi cation
ag
Stream
identi cation
ag
Stream
identi cation
ag
D/O2 0 1 0
D/O3 0 0 1
D/O4 1 1 1
Stream valve number 1 2 2 4 5 5
* Select Up to 7 streams and Starting D/O number for stream identi cation: 4-1 for GCM1. Use one D/O card (5 ch).
* Select Up to 7 streams and Starting D/O number for stream identi cation: 5-1 for GCM2. Use one D/O card (5 ch).
* The starting stream valve number and number of stream valves of GCM1 are “1” and “3,” respectively.
* The starting stream valve number and number of stream valves of GCM2 are “4” and “3,” respectively.
* The above settings are for the case where output exists for calibration (or validation).
In the case where output does not exist for calibration (or validation), the stream identi cation signal of the most recently
measured stream other than calibration or validation is held.
2nd Edition : May 11, 2012-00