Instruction Manual
    <Appendix C Standard Specifi cations>
App.C-7
IM 11B08A01-01E
Power supply: 
100 to 240 V AC ±10%, 
50/60Hz ±5%, 10W or less
Earth: Protective earth
5.2 Signal interrupter
•  For Ethernet twisted pair cables:
  (Rack-mounted type: K9806AA, Desktop 
type: K9806AB)
Number of port:  2
Communication speed:  maximum of 
100Mbps
Power supply:  24 V DC (Supplied from 
External I/O Cutoff Output of CPU 
board in GC8000)
Earth: Functional earth
•  For RS-422 output:  K9806AE
Number of port:  2
Communication speed:  maximum of 
38400bps
Power supply:  24 V DC (Supplied from 
External I/O Cutoff Output of Serial 
communication card in GC8000) 
Earth: Functional earth
•  For analog input:  K9806AE
Number of input:  4
Input current:  4 to 20 mA DC
Input voltage: 1 to 5 V DC
Power supply:  24 V DC (Supplied from 
External I/O Cutoff Output of Analog 
input card in GC8000) 
Earth: Functional earth
•  For contact output (AC):  K9806AN
Number of output:  5
Rated input:  240 V AC, 2A DC
Power supply:  24 V DC (Supplied from 
External I/O Cutoff Output of Contact 
output card in GC8000)
Earth: Protective earth
•  For contact output (DC):  K9806AJ
Number of output:  5
Rated input:  30 V DC, 2A DC
Power supply:  24 V DC (Supplied from 
External I/O Cutoff Output of Contact 
output card in GC8000)
Earth: Functional earth
2nd Edition : May 11, 2012-00
App.C