Instruction Manual
    <Appendix C Standard Specifi cations>
App.C-1
IM 11B08A01-01E
Appendix C  Standard Specifi cations
1. General specifi cations
Measurable object:
  Gas or volatile liquid (400°C or lower 
boiling point)
Analysis method:
 Gas chromatography
Detector: TCD (thermal conductivity detector)
 high-sensitivity TCD
 FID (fl ame ionization detector)
  FID with methanizer
 FPD (fl ame photometric detector) (only 
for large isothermal oven)
Number of detectors:
Type 1:  Maximum of 2
  Selectable from TCD: 2, FID: 2, 
FID with methanizer: 1, FPD: 1
Type 2:  Maximum of 4
  Selectable from TCD: 4, FID: 4, 
FID with methanizer: 2, FPD: 1
Type 3:  Maximum of 6
  Selectable from TCD: 6, FID: 6, 
FID with methanizer: 3
Measurable range:
  Depends on analysis conditions
TCD:  1 ppm to 100%
FID:  1 ppm to 100%
FID with methanizer: 1 ppm to 0.1%
FPD:  1 ppm to 0.1%
Number of components to be measured:
  Maximum of 999 (total number of 
components in all streams including 
calibration standard sample streams)
Number of streams to be measured:
  Maximum of 31 (including calibration 
standard sample streams)
Air output for automatic stream switching:
  Automatic stream valves can be 
directly operated for up to eight 
streams;
  Code switching circuits are 
necessary for nine or more streams.
Note:  Applicable only to 1GCM
Contact output for automatic stream switching:
  Maximum of 20 points
Analysis period:
  Maximum of 21600.0 seconds (six 
hours)
Quantifying method:
  Absolute calibration, sensitivity-
corrected absolute calibration, and 
corrected area normalization
Utility gas supply method:
  EPC (electronic pressure controller) or 
mechanical pressure regulator
  Maximum of 18 streams (up to 6 for an 
isothermal oven)
Number of valves:
Type 1:  Up to 8
  Selectable from 7 Rotary Valves 
(RV), 1 Liquid-sample Valve (LSV), 
and 2 Atmospheric Balance Valves 
(ATM-V)
  Number of RV and LSV is 7 or less
Type2:  Up to 15
  Selectable from 12 RV, 2 LSV, 4 
ATM-V
  Number of RV and LSV is 12 or less
Type3:  Up to 21
  Selectable from 15 RV, 3 LSV, 6 
ATM-V
  Number of RV and LSV is 15 or less
Note:  RVs are not available as Liquid-sample Valve, for 
ATEX applications.
Material of sample-contact parts:
RV:  316SS, Hastelloy-C, Rulon, PTFE 
(Tefl on, Bearee)
LSV:  316SS, Hastelloy-C, Rulon, Glass, 
PTFE (Tefl on, Bearee), Fluororubber 
(Viton), perfl oroelastomer (Kalrez)
Sampling connection:
  6 mm or 1/4 tube (including calibration 
standard sample streams)
  Maximum of 6 streams (up to 2 for 
each isothermal oven)
Note:  Only one Liquid-sample Valve can be mounted in 
each isothermal oven.
Repeatability: Depends on analysis conditions
Gas sample: ±1% of full scale for measuring 
ranges (2)
Liquid sample: ±2% of full scale for measuring 
ranges (2)
Ambient condition during operation:
  Depends on analysis conditions
–10 to 50°C, 95%RH or less (no 
condensation)
Ambient condition during storage:
  –40 to 85°C, no condensation
Installation location:
  Maximum altitude of 2000 m
  Avoid exposure to wind, rain, sunlight
Type of protection:
  Pressurized enclosure and fl ameproof 
enclosure
Certifi cation standard (organization):
  FM, ATEX (DEKRA), IECEx 
(DEKRA), TIIS
FM:  Type X Purging and Explosionproof 
for Class I, Division 1,Groups B, C 
and D. T1 to T4 (Described as FM-X 
hereafter)
  Type X and Y Purging for Class I, 
Division 1, Groups B, C and D. T1 to 
T4 (Described as FM-Y hereafter)
2nd Edition : May 11, 2012-00
App.C