Instruction Manual
    <7. Troubleshooting>
7-4
IM 11B08A01-01E
No. MESSAGE Alarm Contents Threshold Item Threshold Threshold 
Setting
61 AI A/D ERR AI card A/D 
calibration error
Calibrate 16-bit 
AD of temperature 
controller signal 
and AI signal upon 
power-on
Error to calibrate 16-
bit AD of temperature 
controller signal 
and AI signal upon 
power-on 
N/A
62 TMP CARD1 A/D 
ERR
Temp. ctrl card 1 A/D 
cal. error
63 TMP CARD2 A/D 
ERR
Temp. ctrl card 2 A/D 
cal. error
64 TMP CARD3 A/D 
ERR
Temp. ctrl card 3 A/D 
cal. error
65 OVEN1 TMP A/D 
ERR
Oven 1 temp. A/D 
cal. error
66 OVEN2 TMP A/D 
ERR
Oven 2 temp. A/D 
cal. error
67 OVEN3 TMP A/D 
ERR
Oven 3 temp. A/D 
cal. error
71 OVEN1 COM ERR Oven 1 
communication error
Communication with 
OVEN 1
Communication 
between OVEN and 
CPU is disconnected 
for a set period of 
time
N/A
72 OVEN2 COM ERR Oven 2 
communication error
Communication with 
OVEN 2
73 OVEN3 COM ERR Oven 3 
communication error
Communication with 
OVEN 3
74 TMP1 CTL ERR Temp. sensor 1 
temp. cntrl error 
Exceeds the 
deviation limit 
setting for Oven 
1 temperature 
setting value and 
measurement value.
Deviation exceeds 
the value set
Deviation 
temperature 
setting screen
75 TMP2 CTL ERR Temp. sensor 2 
temp. cntrl error 
Exceeds the 
deviation limit 
setting for Oven 
2 temperature 
setting value and 
measurement value.
76 TMP3 CTL ERR Temp. sensor 3 
temp. cntrl error 
Exceeds the 
deviation limit 
setting for Oven 
3 temperature 
setting value and 
measurement value.
77 TMP1 AD ERR Temp. sensor 1 
standard AD error
Standard ZERO/
MIDDLE/SPAN 
values exceed the 
range
Each value exceeds 
x
N/A
78 TMP2 AD ERR Temp. sensor 2 
standard AD error
79 TMP3 AD ERR Temp. sensor 3 
standard AD error
81 TMP1 OVER LIMIT Temp. sensor 1 over 
limit
Temperature sensor 
value
Upper limit: User-set 
value
Lower limit: –20°C
N/A
82 TMP2 OVER LIMIT Temp. sensor 2 over 
limit
83 TMP3 OVER LIMIT Temp. sensor 3 over 
limit
84 TMP1 HIGH Overtemp protection 
zone1
N/A
85 TMP2 HIGH Overtemp protection 
zone2
86 TMP3 HIGH Overtemp protection 
zone3
87 TMP1 SENSOR 
BURN OUT
Temp. sensor 1 
failure
Temperature sensor 
value
Temperature 
exceeds the upper 
limit of 350°C
N/A
88
TMP2 SENSOR 
BURN OUT
Temp. sensor 2 
failure
89 TMP3 SENSOR 
BURN OUT
Temp. sensor 3 
failure
2nd Edition : May 11, 2012-00