Instruction Manual

<6. Maintenance>
6-56
IM 11B08A01-01E
(4) Close all the stream switching valves.
IMPORTANT
During the purge, purge gas may ow into the sample return line, causing pressure
uctuation.
Prepare the sample return line and collection tank, if necessary.
Prepare the sample return line so that the sample does not return.
CAUTION
The isothermal oven is extremely hot after turning off the power immediately. Keep the purging
air supplied for more than an hour after turning off the power. Keep hands away from the oven
components.
(5) Remove the pipe connected to the elbow on the rear of the pressure gauge.
(6) Remove the two screws on the tting.
(7) Remove two nuts and two washers that x the pressure gauge to the tting.
(8) Remove the elbow connected to the pressure gauge.
(9) Install a new pressure gauge in the reverse order of steps (5) to (8). When attaching the
elbow, always apply sealing tape to the connection of the pressure gauge.
(10) Con rm that there are no leaks.
<Replacement Parts>
L9867AG 0 to 0.2 MPa
L9867AJ 0 to 0.6 MPa
F0248.ai
Pressure gauge
Washer
(two)
Nut
(two)
Elbow
Screw (two)
Bracket
Figure 6.48 Pressure Gauge
(5) Pneumatic Valve for Actuating External Air Supply
See Figure 6.49 for the replacement procedure.
(1) Stop the operation. (See “3.3.4 Stopping operation” for this procedure).
2nd Edition : May 11, 2012-00