Instruction Manual

<6. Maintenance>
6-8
IM 11B08A01-01E
<Replacement Parts>
K9401JG
Figure 6.4 Pressure Switch for Detecting Internal Pressure
(2) Fuses
The fuses discussed here are installed in the power line to the protection system. They are
housed in the container of the protection system A on the right side.
(1) Stop the operation (See “3.3.4 Stopping operation” for this procedure).
(2) Stop the supply of power.
(3) Con rm that the LED (Green) of “POWER” is OFF.
CAUTION
The isothermal oven is extremely hot after turning off the power immediately. Keep the purging
air supplied for more than an hour after turning off the power. Keep hands away from the oven
components.
(4) Loosen the locknut on the cover for the protection system A on the right side and remove the
cover.
(5) The fuses are located on the front side, as shown in Figure 6.5.
(6) Use a athead screwdriver to rotate the fuse holder cap counterclockwise and remove it
from the fuse holder.
(7) Replace the fuse in the fuse holder cap with a new one.
Fuse rating Dimensions: ø5.2 × 20 mm
Rated voltage: 250 V AC
Rated current: 1.25 A
Fusing time: Time-delay fusing type
(8) Con rm that the fuses are securely installed in the cap.
(9) Push the cap into the fuse holder until secured.
(10) Con rm that the cap is rmly installed on the fuse holder.
(11) Install the cover on the protection system. Rotate the cover until it stops, then tighten the
locknut.
<Replacement Parts>
A1423EF
2nd Edition : May 11, 2012-00