Instruction Manual
  <5. EtherLCD>
5-99
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM # GCM Number 
Temp Unit  0) °C 
1) Fahrenheit (Refer to SI unit 
limitation)
Det number of 
SIMDISM
 0) NA 
1) 1-1 
2) 1-2 
3) 2-1 
4) 2-2 
5) 3-1 
6) 3-2
Only detectors assigned to 
relevant GCM can be set up.
Response factor 
updating
 Not done, Done
Section updating  Not done, Done
Calilb curve 
updating
 Not done, Done
  SimDis Calib Data Setup
Press F4 (Curve).
11/11/22
15:15:45
SimDis Calib Data Setup  (1/3)
11/11/22
15:15:45
SimDis Calib Data Setup  (2/3)
GCM # 1  /  SYS # 1  /  Method # 1 GCM # 1  /  SYS # 1  /  Method # 1
> - Data acquisition str  0 > - Data acquisition peak3  0
 - Calib curve coef a  1.000  - Elapsed time 3  0.0  s
 - Calib curve coef b  1.000  - Std B.P 3  0.0
 - Data acquisition peak1  0  - Data acquisition peak4  0
 - Elapsed time 1  0.0  s  - Elapsed time 4  0.0  s
 - Std B.P 1  0.0  - Std B.P 4  0.0
 - Data acquisition peak2  0  - Data acquisition peak5  0
 - Elapsed time 2  0.0  s  - Elapsed time 5  0.0  s
 - Std B.P 2  0.0  - Std B.P 5  0.0
Menu SYS#
Method#
Menu SYS#
Method#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.112  Example of SimDis Calib Data Set (1/3) screen 
F1 (Dist):  Displays the SimDis Setup screen.
F5 (SYS #):  Specifi es a SYS number.
When F5 (SYS #) is pressed, “SYS No.:” is displayed on the second line below. Use the numeric 
keypad to enter the number and press the SET/ENT key, and a cursor jumps to the SYS number. 
Only the SYS numbers set in GCM is selectable. When any other number is entered, “Over limit” 
is displayed on the second line below.
•  F6 (Method #): Specifi es a Method number.
•  The GCM number displayed on the second line above is the same as the one set on the 
SimDis Setup screen.
•  Data acquisition streamSet the cal run stream number acquired for acquiring data to 
compile a SimDis calib curve. On the stream setup screen, when a stream number with 
other stream type than Cal run is about to be set, the error message “No cal run stream” is 
displayed on the second line below.
•  Calib curve coef a & b
  Linear ramp a & slice b of a SimDis Calib curve. Auto updating and manual entry are 
possible.
2nd Edition : May 11, 2012-00