Instruction Manual

<5. EtherLCD>
5-96
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Period # Period script Number 1 64
GCM Number 06
Auto Start * Not done (not executable)
Done (executable)
Exec Interval Date
Hour
Minute
Second
Ana end
Interval (Date) 1 31 Date When not selected by Cycle
Time Unit, please display with
asterisk. However, leave internal
data as they are.
Interval (Hour) 1 148 Hour
Interval
(Minute)
1 1440 Minute
Interval
(Second)
1 3600 Second
Exec SYS
Number
1 6 Only SYS numbers belonging to
relevant GCM can be selected.
When Ana end is not selected
in the Cycle Time Unit, display *
(asterisk).
* When Exec interval is Ana
end, the relevant period
script is executed at the time
of Ana end for the speci ed
SYS number.
*: The settings are valid after rebooting or turning off/on its power.
User Program Delete Screen (Period script)
Press F3 (Delete).
11/11/22
15:15:45
User Program Delete
Period # 1
> - User Program Delete Not deleted
Menu Main Period#
F1 F2 F3 F4 F5 F6
Figure 5.108
Example of User Program Delete (Period script) screen
F1 (Menu): Displays the Table menu screen.
F2 (Main): Displays the User Programming screen (Period script).
F6 (Period #): Speci es a Period script number.
The operation accepts user level above C only.
When Delete is selected in the User Program Delete, the relevant period script le is
deleted. When a script is deleted, the auto start of the script is automatically changed to NA.
Any period script in the status Enable cannot be deleted (an error message is displayed).
When the execution status is Enable and the user program is about to be deleted, the error
message “No execution because of executing” is displayed.
Even when the user program is deleted, the script le in the SD card is not deleted.
However, the script on the user program is deleted.
2nd Edition : May 11, 2012-00