In some SYSs, when a valve is set to Yes and other SYS is selected, **** is displayed, “Already
set in other SYS” is displayed in the second line below and it cannot be set again even though it
is selected.
Any SYS number that does not exist in F5 (SYS#) cannot be entered.
Any oven number that does not exist in F6 (Oven#) cannot be entered.
Detector Setup
Press F5 (SYS#), and “SYS No.” is displayed in the second line below.
Use the numeric keypad to input the number and press the Set/Ent key, and it is re ected in the
second line above and the SYS contents of the number is shown.
Press F6 (Oven#), and “Oven No.” is displayed in the second line below.
Use the numeric keypad to select the number and press the Set/Ent key, and it is re ected in the
second line above, and the valve set of the speci ed oven is completed.
Point a cursor to a detector and press the Set/Ent key, and options are displayed in the second
line below.
Select either Yes or NA.
Not installed detectors are displayed as ----.
When any of the Not installed detectors is used by other SYS, **** is displayed.
In some SYSs, when a detector is set to Yes and other SYS is selected, **** is displayed,
“Already set in other SYS” is displayed in the second line below and it cannot be set again even
though it is selected.
EPC Setup
Press F5 (SYS#), and “SYS No.:” is displayed in the second line below.
Use the numeric keypad to select the number and press the Set/Ent key, and the SYS contents
of the number is shown.
Press F6 (Oven#), and “Oven No.:” is displayed in the second line below.
Use the numeric keypad to select the number and press the Set/Ent key, and it is re ected in the
second line above and the valve set of the speci ed oven is completed.
Point a cursor to EPC and press the Set/Ent key, and options are displayed in the second line
below.
Point a cursor to an item and press the Set/Ent key, and options are displayed in the second line
below. Select either Yes or NA and press the Set/Ent key, and the setting is done.
When Yes is selected for either EPC: CarrierN-1 or CarrierN-2 that is set in the Program on the
EPC setup screen, EPC program set in the SYS method set is available.
(When Yess are selected for EPC: CarrierN-1, CarrierN-2, and Utility N-1 to N-4 that are set to
Constant in the EPC setup screen Yes, enter set values in the Operating Parameters.)
Not installed EPCs are displayed as ----.
When any of the Not installed EPC is used by other SYS, **** is displayed.
In some SYSs, when an EPC is set to Yes and other SYS is selected, **** is displayed, “Already
set in other SYS” is displayed in the second line below and it cannot be set again even though it
is selected.
Up to 2 carrier gases set in the program can be set for one SYS.
Oven Setup
Press F5 (SYS#) and “SYS No.:” is displayed in the second line below.
Use the numeric keypad to select the number and press the Set/Ent key, and the SYS contents
of the number is shown.
2nd Edition : May 11, 2012-00