F1 (Menu): Displays the Table menu screen.
F3 (Status): Displays the A/O status screen.
F5 (Slot #): Sets a Slot number.
F6 (AO #): Sets a A/O number.
When A/O chromatogram output is set, other items are displayed with * (asterisk).
#1 Output stream and #2 Output stream cannot be set in the same Stream. If this setup is
attempted, the error message “Already assigned in request stream” is displayed.
If a detector that does not exist is selected in the setting of the A/O chromatogram, the error
message “Not load the detector” is displayed.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Slot# Slot number 1 5
AO # A/O channel number 1 8
GCMNumber 16
AO
Chromatogram
Normal output, Det1-1, Det1-2,
Det2-1, Det2-2, Det3-1, Det3-2
* Multiple OA channels can
be assigned to the same
detector. However, up to a
maximum of 8 channels.
#1 Output
stream
(0), 1 (99), 31 Lower Limit (0) cancels the
output of the relevant peak.
Upper Limit (99) gives output of
analyzed values of all streams.
#2 Output
stream
(0), 1 31 Lower Limit (0) cancels the
output of the relevant peak.
#1, #2 Output
peak
(0), 1 999 Lower Limit (0) cancels the
output of the relevant peak.
#1, #2 Partial
hi
0.001 1.000
#1, #2 Partial
low
0.000 0.999
5.4.18 A/I Setup
11/11/22
15:15:45
A/I Setup
Slot # 1 AI # 1
> - Signal name AA
- Filter const 1.000
Menu Status Slot# AI#
F1 F2 F3 F4 F5 F6
Figure 5.99
Example of A/I setup screen
F1 (Menu): Displays the Table menu screen.
F3 (Status): Displays the A/I status screen.
F5 (Slot#): Sets a Slot number.
F6 (AI#): Sets an A/I number.
2nd Edition : May 11, 2012-00