Instruction Manual
<5. EtherLCD>
5-86
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Slot # 1 5 Limited
DI # 1 8 Limited
Signal name Alphanumeric: 8 characters
Process
assign
0: No process, 1: Stream
sequence, 2: Stream, 3: Range
select, 4: Cal/Val, 5: Operation
mode, 6: Alarm process
GCM number Setup possible when 0 to 6,
process assign entails Stream
sequence, Cal (Val) and
Operation mode.
* In the case of zero, applicable
to all GCM. When process
assign entails alarm, applicable
to alarms for all GCM and
equipment.
Process
sequence
18
Stream
number
131
Number of
times
0 999 0: Consecutive, 1 to 999: times
Range
assigned
Stream
131
Range
assigned
Peak
1 999
Range
number
131
Cal/Val
number
Cal 1, Cal 2, Cal 3, Cal 4, Cal 5,
Cal 6, Val 1, Val 2, Val 3, Val 4,
Val 5, Val 6
Operation
mode
0: Run, 1: Stop, 2: Pause
Alarm level 2 or 3
Alarm number 201 to 232 (in the case of alarm
level 2), 401 to 432 (in the case
of alarm level 3)
Message Alphanumeric: 22 characters
5.4.17 A/O Setup
When #1 Output stream is set at 99, the analyzed value of all streams belonging to GCM are
read out. In this case, the setup for #2 Output stream and #2 Output peak become invalid. This
setup can be done independently for each A/O.
11/11/22
15:15:45
A/O Setup (1/2)
11/11/22
15:15:45
A/O Setup (2/2)
Slot # 1 AO # 1 Slot # 1 AO # 1
> - GCM number 1 > - #2 Output stream * *
- AO Chromatogram Det 1-1 #2 Output peak * * *
- #1 Output stream * * - #2 Percent span offset * * * * *
#1 Output peak * * * - #2 Percent zero offset * * * * *
- #1 Percent span offset * * * * *
- #1 Percent zero offset * * * * *
Menu Status Slot# AO# Menu Status Slot# AO#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.98
Example of A/O setup screen
2nd Edition : May 11, 2012-00