Instruction Manual
  <5. EtherLCD>
5-86
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Slot # 1 5 Limited
DI # 1 8 Limited
Signal name  Alphanumeric: 8 characters
Process 
assign
 0: No process, 1: Stream 
sequence, 2: Stream, 3: Range 
select, 4: Cal/Val, 5: Operation 
mode, 6: Alarm process
GCM number   Setup possible when 0 to 6, 
process assign entails Stream 
sequence, Cal (Val) and 
Operation mode. 
* In the case of zero, applicable 
to all GCM. When process 
assign entails alarm, applicable 
to alarms for all GCM and 
equipment.
Process 
sequence
 18
Stream 
number 
 131
Number of 
times
 0 999 0: Consecutive, 1 to 999: times
Range 
assigned 
Stream 
 131
Range 
assigned 
Peak 
 1 999
Range 
number 
 131
Cal/Val 
number 
 Cal 1, Cal 2, Cal 3, Cal 4, Cal 5, 
Cal 6, Val 1, Val 2, Val 3, Val 4, 
Val 5, Val 6
Operation 
mode
 0: Run, 1: Stop, 2: Pause
Alarm level   2 or 3
Alarm number   201 to 232 (in the case of alarm 
level 2), 401 to 432 (in the case 
of alarm level 3)
Message Alphanumeric: 22 characters
5.4.17 A/O Setup
When #1 Output stream is set at 99, the analyzed value of all streams belonging to GCM are 
read out. In this case, the setup for #2 Output stream and #2 Output peak become invalid. This 
setup can be done independently for each A/O. 
11/11/22
15:15:45
A/O Setup  (1/2)
11/11/22
15:15:45
A/O Setup  (2/2)
Slot # 1  AO # 1 Slot # 1  AO # 1
> - GCM number  1 > - #2 Output stream  * *
 - AO Chromatogram  Det 1-1  #2 Output peak  * * *
 - #1 Output stream  * *  - #2 Percent span offset  * * * * *
 #1 Output peak  * * *  - #2 Percent zero offset  * * * * *
 - #1 Percent span offset  * * * * *
 - #1 Percent zero offset  * * * * *
Menu Status Slot# AO# Menu Status Slot# AO#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.98 
Example of A/O setup screen
2nd Edition : May 11, 2012-00