Instruction Manual
<5. EtherLCD>
5-83
IM 11B08A01-01E
11/11/22
15:15:45
D/O Setup
Slot # 1 DO # 1
> - Process assign Timing
- SYS number 1 (GCM 1)
- SYS Method 1
- First time AAAAAA
- Second time AAAAAA
- Third time AAAAAA
Menu Status Slot# DO#
F1 F2 F3 F4 F5 F6
Figure 5.94
Example of D/O setup screen (setting the Process assign to Timing)
• Any SYS number that is not set in GCM cannot be set.
• When an Over limit value is set as a SYS number, “Over limit” is displayed in the bottom line
and the set value is not shown.
11/11/22
15:15:45
D/O Setup (1/2)
11/11/22
15:15:45
D/O Setup (2/2)
Slot # 1 DO # 1 Slot # 1 DO # 1
> - Process assign Cal/Val > - Validation 1 AAAA
- GCM number 1 - Validation 2 AAAA
- Calibration 1 AAAA - Validation 3 AAAA
- Calibration 2 AAAA - Validation 4 AAAA
- Calibration 3 AAAA - Validation 5 AAAA
- Calibration 4 AAAA - Validation 6 AAAA
- Calibration 5 AAAA
- Calibration 6 AAAA
Menu Status Slot# DO# Menu Status Slot# DO#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.95
Example of D/O setup screen (setting the Process assign to Cal/Val)
• When an Over limit value is set as a GCM number, “Over limit” is displayed in the bottom
line and the set value is not shown.
11/11/22
15:15:45
D/O Setup
Slot # 1 DO # 1
> - Process assign Str valve select
- Str valve number 1
Menu Status Slot# DO#
F1 F2 F3 F4 F5 F6
Figure 5.96
Example of D/O setup screen (setting the Process assign to Str valve select)
• The Str valve select is a function to operate stream selection valves, and is operated
according to the Stream valve ON/OFF time in the Method.
2nd Edition : May 11, 2012-00
5