D/O Setup 
Slot # 1     DO # 1
> - Process assign  Timing
 - SYS number  1 (GCM 1)
 - SYS Method  1
 - First time  AAAAAA
 - Second time  AAAAAA
 - Third time  AAAAAA
Menu Status Slot# DO#
F1 F2 F3 F4 F5 F6
Figure 5.94 
Example of D/O setup screen (setting the Process assign to Timing)
•  Any SYS number that is not set in GCM cannot be set.
•  When an Over limit value is set as a SYS number, “Over limit” is displayed in the bottom line 
and the set value is not shown.
11/11/22
15:15:45
D/O Setup  (1/2)
11/11/22
15:15:45
D/O Setup  (2/2)
Slot # 1  DO # 1 Slot # 1  DO # 1
> - Process assign  Cal/Val > - Validation 1  AAAA
 - GCM number  1  - Validation 2  AAAA
 - Calibration 1  AAAA  - Validation 3  AAAA
 - Calibration 2  AAAA  - Validation 4  AAAA
 - Calibration 3  AAAA  - Validation 5  AAAA
 - Calibration 4  AAAA  - Validation 6  AAAA
 - Calibration 5  AAAA
 - Calibration 6  AAAA
Menu Status Slot# DO# Menu Status Slot# DO#
F1 F2 F3 F4 F5 F6 F1 F2 F3 F4 F5 F6
Figure 5.95 
Example of D/O setup screen (setting the Process assign to Cal/Val)
•  When an Over limit value is set as a GCM number, “Over limit” is displayed in the bottom 
line and the set value is not shown.
11/11/22
15:15:45
D/O Setup 
Slot # 1     DO # 1
> - Process assign  Str valve select
 - Str valve number  1
Menu Status Slot# DO#
F1 F2 F3 F4 F5 F6
Figure 5.96 
Example of D/O setup screen (setting the Process assign to Str valve select)
•  The Str valve select is a function to operate stream selection valves, and is operated 
according to the Stream valve ON/OFF time in the Method.