Instruction Manual

<5. EtherLCD>
5-80
IM 11B08A01-01E
5.4.14 Communications Setup
11/11/22
15:15:45
Communications Setup
COM # 1
> - Type DCS
- Protocol MODBUS
- Trans contents * * * * * * * * * *
- Recep contents * * * * * * * *
- Trans speed 9600bps
- Procedure * * * * * * * * * * * * * * * * *
- Parity Odd
- Com mode RTU mode
- Scaling coef Real
Menu COM#
F1 F2 F3 F4 F5 F6
Figure 5.88
Examples of communications setup screen
F1 (Menu): Displays the Table menu screen.
F6 (COM #): Sets a communication port number.
Modbus protocol can be set to both COM1 and COM2. However, GCCU and GCCU or
Y-Protocol (GC1000/GC8, GC6, GCHC) and Y-Protocol cannot be set to COM1 and COM2
respectively.
Items that cannot be set due to types and protocols are displayed with * (asterisk).
Type NA DCS GCCU
Protocol × MODBUS GC1000/GC8 GC6 GCHC ×
Trans contents × × 
Recep contents × × 
Trans speed × 
Procedure × × 
Parity × 
Com mode × ××××
Scaling coef × ××××
For the Y-Protocol, refer to “ Serial Communication (Y-Protocol)” of page App.D-21
Display Item Name Set Lower Limit Upper Limit Unit Remarks
COM # 1 2 None, DCS, GCCU
Type * None, DCS, GCCU
Protocol * MODBUS, GC1000/GC8, GC6,
GCHC * Only when the type is
DCS
Trans
contents
* Analyzed value, Alarm, All data
Recep
contents
* None, Command, All data, GC8
type
Trans speed * 1200 bps, 2400 bps, 4800 bps,
9600 bps, 19200 bps, 38400
bps
Procedure * Without procedure, With
procedure
Parity * None, Odd, Even
Com mode * RTU mode, ASCII mode
Scaling coef * Real, 9999, 65535
*: The settings are valid after rebooting or turning off/on its power.
2nd Edition : May 11, 2012-00