Instruction Manual
  <5. EtherLCD>
5-79
IM 11B08A01-01E
5.4.13 Peak Assignments 
11/11/22
15:15:45
Peak Assigments  (1/4)
 Total assign 30  unassign 969
Strm# Name  Assign  Use  GCM num
 1  A1  10  0  1
 2  A2  10  5  1
 3  10  5  1
 4  0  0  1
 5  0  0  1
 6  0  0  1
 7  0  0  1
 8  0  0  1
Menu Exec Assign
F1 F2 F3 F4 F5 F6
Figure 5.87 
Example of peak assignments (1/4) screen
F1 (Menu):  Displays the Table menu screen.
F4 (Exec):  Executes confi guration change of peak table.
F6 (Assign):  Sets stream numbers and assign numbers for the setting.
The setting is enabled only when all GCMs are in stop mode.
The total usage of peak table is set per stream.
When a stream set in GCM is selected, the GCM number is displayed in the GCM number area.
When a stream that is not set in GCM is selected, the GCM number area is blank.
Stream numbers and relative peak numbers are reset by executing F4 (Exec). Execution is 
initialized to Off.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Total assign Total assigned peak 0 999
Unassigned Unassigned peak 
number 
0 999
Stream # Stream number  1 31
Stream name If any stream name or set name 
exists, the table content is 
directly shown.
Assignment 
number
Relevant stream 
peak assign number
0 999 Settable
Usage Relevant stream 
peak usage
0 999
GCM Number  1 6
2nd Edition : May 11, 2012-00
5