Instruction Manual

<5. EtherLCD>
5-78
IM 11B08A01-01E
5.4.12 Alarm Setup
11/11/22
15:15:45
Alarm Setup
GCM # 1 Set # 1
> - Stream number 2
Peak number 1
- Check item Conc
- Maximum 0.000
- Minimum 0.000
Menu GCM# Set#
F1 F2 F3 F4 F5 F6
Figure 5.86
Example of alarm setup screen
F1 (Menu): Displays the Table Menu screen.
F5 (GCM #): Speci es a GCM number.
F6 (Set #): Sets an alarm number.
When a stream and a peak number in database are both zero, they are assessed as no
settings and all data is displayed with * (asterisk).
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM #
Set # Relevant set number (1 to 32),
32 per GCM (Total 192)
Stream
number
(0), 1 31 99 can also be entered.
Peak number (0), 1 999 When a stream number is 99,
all streams belonging to GCM is
applicable and a relative peak
number is assigned to the peak
number.
Assign 0 to the stream number
and the peak number to enable
setting invalid data.
Check items Alarm check items, Density, R.
Time, Vari coeff, Tailing coeff
Upper Limit,
Lower Limit
Density: 999.999 to 9999.999
(Process assign is A/I signal),
0.000 to 9999.999 (Process
assign is other than A/I signal)
R. Time:0.000 to 21600.0 sec
Vari coeff: 0.000 to 1.999
Tailing coeff: 0.000 to100.000
2nd Edition : May 11, 2012-00