Instruction Manual
  <5. EtherLCD>
5-78
IM 11B08A01-01E
5.4.12 Alarm Setup
11/11/22
15:15:45
Alarm Setup 
GCM # 1   Set # 1
> - Stream number  2
 Peak number  1
 - Check item  Conc
 - Maximum  0.000
 - Minimum  0.000
Menu GCM# Set#
F1 F2 F3 F4 F5 F6
Figure 5.86 
Example of alarm setup screen
F1 (Menu):  Displays the Table Menu screen.
F5 (GCM #):  Specifi es a GCM number.
F6 (Set #):  Sets an alarm number.
•  When a stream and a peak number in database are both zero, they are assessed as no 
settings and all data is displayed with * (asterisk).
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM #
Set # Relevant set number (1 to 32), 
32 per GCM (Total 192)
Stream 
number 
 (0), 1 31 99 can also be entered.
Peak number  (0), 1 999 When a stream number is 99, 
all streams belonging to GCM is 
applicable and a relative peak 
number is assigned to the peak 
number. 
Assign 0 to the stream number 
and the peak number to enable 
setting invalid data.
Check items  Alarm check items, Density, R. 
Time, Vari coeff, Tailing coeff
Upper Limit, 
Lower Limit
 Density: 999.999 to 9999.999 
(Process assign is A/I signal), 
0.000 to 9999.999 (Process 
assign is other than A/I signal) 
R. Time:0.000 to 21600.0 sec
Vari coeff: 0.000 to 1.999
Tailing coeff: 0.000 to100.000
2nd Edition : May 11, 2012-00