Instruction Manual

<5. EtherLCD>
5-76
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Peak # Peak number 1 999
Peak name Alphanumeric: 8 characters
Execution Executed
Not executed
Measurement
unit
1 100 1 ( )
2 (%)
3 (ppm)
4 (ppb)
5 (wt.%)
6 (wt.ppm)
7 (wt.ppb)
8 (vol.%)
9 (vol.ppm)
10 (vol.ppb)
11 (mol.%)
12 (mol.ppm)
13 (mol.ppb)
14 and over ( )
Measuring
range
0.000 9999.999
Std conc 0.000 9999.999
Standard
area
0.000 9999.999
Calibration
factor
0.000 9999.999
Validation Set
Press F5 (Val).
11/11/22
15:15:45
Validation Setup
GCM # 1 Val # 1
> - Val stream 3
- Val times 1
- Auto val Unexecuted
Menu Main GCM# Val#
F1 F2 F3 F4 F5 F6
Figure 5.84
Example of validation set screen
F1 (Menu): Displays the Table menu screen.
F4(Main): Displays the Cal/Val Set (main) screen.
F5 (GCM #): Speci es a GCM number.
F6 (Val #): Speci es a Val number.
Validation stream can only be set when stream type is Val. When stream type is other than
Val, the message “The kind of stream is incorrect” is displayed. When it does not exist in the
relevant GCM, the message “The stream doesn’t exist”in GCM is displayed
2nd Edition : May 11, 2012-00