Instruction Manual

<5. EtherLCD>
5-75
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM #
Cal # Cal Number 0 6
Cal stream 0 31* * Only stream numbers and cal
streams belonging to GCM can
be set. (Stream setup screen )
Cal times 0 30 Cal times + Val before cal times
+ Val after cal times must not
exceed 31.
Auto cal
speci cation
Not done, Done
ValStr #
before and
after cal
0 31** ** Only stream numbers and Val
streams belonging to GCM can
be set. (Stream setup screen)
Val before
and after cal
times
0 30 Cal times + Val before cal times
+ Val after cal times must not
exceed 31.
Calibration peak set
Press F2 (Peak).
11/11/22
15:15:45
Calibration Peak Setup
Peak # 1
> - Peak name
- Execution Executed
- Measuring unit 1 ( )
- Measuring range 1.000
- Std conc 1.000
- Std area 950.000
- Calib factor 2.063
Cal Peak#
F1 F2 F3 F4 F5 F6
Figure 5.83
Example of calibration peak set screen
F2 (Cal): Displays the calibration set screen
F6 (Peak #): Speci es a peak number.
Displays the stream peak set in Cal stream.
2nd Edition : May 11, 2012-00
5