Instruction Manual
  <5. EtherLCD>
5-75
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM #
Cal # Cal Number  0 6
Cal stream   0 31* * Only stream numbers and cal 
streams belonging to GCM can 
be set. (Stream setup screen )
Cal times  0 30 Cal times + Val before cal times 
+ Val after cal times must not 
exceed 31. 
Auto cal 
specifi cation
 Not done, Done
ValStr # 
before and 
after cal 
 0 31** ** Only stream numbers and Val 
streams belonging to GCM can 
be set. (Stream setup screen)
Val before 
and after cal 
times
 0 30 Cal times + Val before cal times 
+ Val after cal times must not 
exceed 31.
 Calibration peak set
Press F2 (Peak).
11/11/22
15:15:45
Calibration Peak Setup 
Peak # 1
> - Peak name  
 - Execution  Executed
 - Measuring unit  1  (  )
 - Measuring range  1.000
 - Std conc  1.000
 - Std area  950.000
 - Calib factor  2.063
Cal Peak#
F1 F2 F3 F4 F5 F6
Figure 5.83 
Example of calibration peak set screen
F2 (Cal):  Displays the calibration set screen
F6 (Peak #):  Specifi es a peak number. 
•  Displays the stream peak set in Cal stream.
2nd Edition : May 11, 2012-00
5