Instruction Manual
  <5. EtherLCD>
5-74
IM 11B08A01-01E
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM #
Cal (Val) 
method
 Manual, Semi-auto, Auto
Auto start 
date
 2000 to 2099, Jan to Dec, 1st 
to 31st*
* Upper Limit will differ 
depending on the month 
Auto start 
time
 0 to 23 hours, 0 to 59 minutes
Time interval  0 to 255 days, 0 to 23 hours, 0 
to 59 minutes
Date, Hour, and Min cannot be 
set to all 0. 
Auto cal 
status
Stopping, Executing
 Calibration set
Press F4 (Cal).
11/11/22
15:15:45
Calibration Setup 
GCM # 1  Cal # 1
> - Cal stream  2
 - Cal times  1
 - Auto cal  Unexecuted
 - ValStr# before Cal  0
 - ValStr# after Cal  0
 - Val before Cal times  1
 - Val after Cal times  1
Menu Peak Main GCM# Cal#
F1 F2 F3 F4 F5 F6
Figure 5.82 
Example of calibration set screen
F1 (Menu):  Displays the Table Menu screen.
F2 (Peak):  Displays the calibration peak set screen.
F4(Main):  Displays the Cal/Val Set (main) screen.
F5 (GCM #):  Specifi es a GCM number.
F6 (Cal #):  Specifi es a Cal number.
When ValStr # before Cal and ValStr # after Cal are set to 0, Val stream is not executed.
•  When a GCM number is changed, the Cal number is allowed to return to 1.
•  A Cal stream can be set to Cal only when a stream type is set to Cal. When the setting is 
attempted with a stream type other than Cal, the message “The kind of stream is incorrect” 
is displayed. When it does not exist in the relevant GCM, the message “The stream doesn’t 
exist in GCM” is displayed.
•  A ValStr # before Cal and a ValStr # after Cal can only be set when the stream type is Val. 
When the stream type is other than Val, the message “The kind of stream is incorrect” is 
displayed. When it does not exist in the relevant GCM, the message “The stream doesn’t 
exist in GCM” is displayed.
2nd Edition : May 11, 2012-00