Instruction Manual
  <5. EtherLCD>
5-69
IM 11B08A01-01E
Table 5.3  Synchronization rules table 
Destination
Source
Measurement Cal  Val
Calibration 
run
Sample run
Measurement   ×  ××
Cal   ×  ××
Val  ×  ××
Calibration run ×××××
Sample run ××××
, :  Synchronization possible, see Synchronize items for synchronization items. 
 *: The  symbol in the Synchronization Rules Table signifi es (Cal Measurement, Cal Val) and he synchronization items  
(process assignment, reference peak, etc.) cannot be synchronized.
×:  Cannot be synchronized.
The following is an example of setting error.
Synchronized destination: Stream 2ーPeak 1
Stream 1ーPeak 2
F0502.ai
Stream 1ーPeak 1
Indirect synchronization
Contradicts the rule “Cannot set peak within the same stream to synchronized destination.”
2nd Edition : May 11, 2012-00
5