Instruction Manual
<5. EtherLCD>
5-63
IM 11B08A01-01E
5.4.8 Peak Setup-General
11/11/22
15:15:45
Peak Setup-General
GCM #1
> - Conc calc req Corr area norm
- Det slope ratio 0.00
- Calib range 0.500
- Calib repeat range 2.000
- Calib coef range 4.000
- #1 Tickmark setup None
- #2 Tickmark setup None
- Peak detect err set Executed
Menu GCM#
F1 F2 F3 F4 F5 F6
Figure 5.77
Example of the peak setup-general screen
F1(Menu): Displays Table Menu screen
F6(GCM#): Specifi es GCM number.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM # 1 6 0: Standard, 1: Correction area
fraction
Conc cal req
Det slope
ratio
0.00 9.99
Calib range 0.000 1.000
Calib repeat
range
0.000 2.000
Calib coef
range
0.000 4.000
#1 Tickmark
setup
None, Start, 10s, Start 10s,
60s, Start 60s, 10s to 60s,
Start 10s to 60s
#2 Tickmark
setup
None, Gate, Peak, Gate to Peak
Peak detect
err set
Not executed, Executed
2nd Edition : May 11, 2012-00