Instruction Manual
  <5. EtherLCD>
5-61
IM 11B08A01-01E
5.4.6 Stream Setup
11/11/22
15:15:45
Stream Setup 
Stream # 1
> - Name of stream  
 - Str valve number  0
 - Type of stream  Measurement
 - Normalization  * * * * * * * * * *
 - Total value  100.0 %
 - Base line corr  Unexecuted
 - GCM number  1
 - Method number  1
  - SIMDIS calib stream  
Menu Strm#
F1 F2 F3 F4 F5 F6
Figure 5.75 
Example of the stream setup screen
F1 (Menu):  Displays Table Menu screen.
F6 (Strm #):  Specifi es a stream number (1 to 31).
•  The SimDis calib stream item is displayed only when distillation is set to Yes in the GCM 
setup (2/2) screen for the GCM specifi ed by the GCM number. Displays a stream type when 
it is a sample run. 
•  In the case of the area correction percentage for GCM, normalization for the stream 
belonging to that GCM is displayed by an asterisk.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Stream # Stream number  1 31
Stream name   Alphanumeric: 8 characters
Stream valve 
number 
 0 31* 0: No stream valve
* Only stream valve number 
belonging to the GCM can be 
set (GCM setup screen).
Stream type   Measurement, Cal, Val, 
Calibration run (only when 
distilling), Sample run (only 
when distilling)
Normalization   Not executed, Executed
* When the GCM of the 
relevant stream is confi gured 
with multiple SYS, Executed 
cannot be set.
Total value   0.1 100.0
Base line corr   Not executed, Executed
GCM number   16
Method 
number 
 16
Distillation 
confi guration 
stream
 031
2nd Edition : May 11, 2012-00