Instruction Manual

<5. EtherLCD>
5-61
IM 11B08A01-01E
5.4.6 Stream Setup
11/11/22
15:15:45
Stream Setup
Stream # 1
> - Name of stream
- Str valve number 0
- Type of stream Measurement
- Normalization * * * * * * * * * *
- Total value 100.0 %
- Base line corr Unexecuted
- GCM number 1
- Method number 1
- SIMDIS calib stream
Menu Strm#
F1 F2 F3 F4 F5 F6
Figure 5.75
Example of the stream setup screen
F1 (Menu): Displays Table Menu screen.
F6 (Strm #): Speci es a stream number (1 to 31).
The SimDis calib stream item is displayed only when distillation is set to Yes in the GCM
setup (2/2) screen for the GCM speci ed by the GCM number. Displays a stream type when
it is a sample run.
In the case of the area correction percentage for GCM, normalization for the stream
belonging to that GCM is displayed by an asterisk.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
Stream # Stream number 1 31
Stream name Alphanumeric: 8 characters
Stream valve
number
0 31* 0: No stream valve
* Only stream valve number
belonging to the GCM can be
set (GCM setup screen).
Stream type Measurement, Cal, Val,
Calibration run (only when
distilling), Sample run (only
when distilling)
Normalization Not executed, Executed
* When the GCM of the
relevant stream is con gured
with multiple SYS, Executed
cannot be set.
Total value 0.1 100.0
Base line corr Not executed, Executed
GCM number 16
Method
number
16
Distillation
con guration
stream
031
2nd Edition : May 11, 2012-00