Instruction Manual
  <5. EtherLCD>
5-58
IM 11B08A01-01E
•  Only the peak belonging to the relevant SYS can be selected. Default displays the peak with 
the smallest number belonging to the SYS.
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM # GCM number  1 6
SYS # SYS number  1 6
Method # Method number  1 6
Stream # Stream number  1 31
Peak number  Peak number  1 Assigned 
peak number 
Default displays peak number 1.
After that, the previous peak 
number is displayed.
Gate ON time Gate ON time  0.0 *1
21600.0
seconds  Some data are invalid.
*1: Upper limit varies depending 
on the sample rate of the 
detector number set for the 
relevant peak.
Sample rate Max
20 ms  2700.0 s
40 ms 5400.0 s 
80 ms 10800.0 s
160 ms 21600.0 s
Gate OFF 
time
Gate OFF time 
  Valve Action Setting Screen
Press F3 (Valve).
11/11/22
15:15:45
Valve Action Setting 
GCM #1  /  SYS #1  /  Method #1
Valve number    1-1
> - 1st ON  time  0.0 s
 - 1st OFF time  5.0 s
 - 2nd ON  time   * * * * * * * s
 - 2nd OFF time   * * * * * * * s
 - 3rd ON  time   * * * * * * * s
 - 3rd OFF time   * * * * * * * s
Seq Gate AtmV DO Valve#
F1 F2 F3 F4 F5 F6
Figure 5.72 
Example of the valve action setting screen
F1(Seq):  Displays SYS Sequential Display screen.
F2(Gate):  Displays Gate screen of the action set screen
F4(Atm):  Displays ON/OFF time screen of ATM Valve (1st, 2nd & 3rd).
F5(DO):  Displays D/O Setup screen in the action set screen.
F6(Valve#): Specifi es a using valve number.
•  Sets possible for User Level C and above.
Pressing F6 (Valve #) displays Oven number in the second line on the bottom. 
After selecting an oven number, Valve number is displayed. 
Selecting the valve number and pressing ENT sets the valve ON/OFF time of that number.
Executing ENT with a cursor pointing at each set displays the cursor in the second line on the 
bottom.
2nd Edition : May 11, 2012-00