Instruction Manual
<5. EtherLCD>
5-57
IM 11B08A01-01E
F1(Main): Displays SYS Method Setup screen (Main).
F2(Gate): Displays Action Set (gate) in the action set screen.
F3(Valve): Displays ON/OFF time screen in a valve (1st, 2nd & 3rd of the maximum No. 1
to No. 18).
F4(Atm): Displays ON/OFF time screen of ATM Valve (1st, 2nd & 3rd).
F5(DO): Displays D/O Setup screen in the action set screen.
F6(Strm#): Specifi es a stream number.
• Stream number is assigned with F6 (Stm #). Only the stream number being used by the
relevant GCM number can be set. The smallest stream number is displayed in the initial
display.
• Up to 512 sequences can be displayed.
• SYS sequence with SYS usage set as “Not executed” is also displayed.
• Gate ON/OFF time for the peak not executed is not displayed.
• Base/signal/noise level and A/I signals 1 to 4 of the action time are displayed.
• When D/O is not installed, F5 (DO) is not displayed.
• When any peak is not assigned, selecting F2 (gate) displays the message “No assignment
peak of request tream.”
Display Item Name Set Lower Limit Upper Limit Unit Remarks
GCM # GCM number 1 6
SYS # SYS number 1 6
Method # Method number 1 6
Stream # Stream number 1 31
Time
Action
Gate Action Setting Screen
Press F2 (Gate).
11/11/22
15:15:45
Gate Action Setting
GCM #1 / SYS #1 / Method #1 / Stream #1
Peak number 1
> - Gate ON time 1.0 s
- Gate OFF time 5.0 s
Main Valve AtmV DO Peak#
F1 F2 F3 F4 F5 F6
Figure 5.71
Example of the gate action setting screen
F1(Seq): Displays SYS Sequential Display screen.
F3(Valve): Displays Valve screen of the action set screen.
F4(Atm): Displays ON/OFF time screen of ATM Valve (1st, 2nd & 3rd).
F5(DO): Displays D/O Setup screen in the action set screen.
F6(Peak#): Specifi es a peak number (relative peak number).
• Sets possible for User Level C and above.
2nd Edition : May 11, 2012-00
5